12
Trigger VNA to perform a single sweep.
Figure 65 Setup to measure "Vout" of probe
InfiniiMax III Series Probes User's Guide
Press "Trigger" under Channel Setup, and then the green soft-key for
a
"Single".
Under "Scale" menu, adjust the reference level until the 50 MHz point (left
b
side of the screen) is at center screen. Reference level should be
approximately -15.3 dB, but can vary a few tenths of a dB either way.
Display should look like
c
tip wires may not be making contact under the pinchers.
Figure
66. If it looks noticeably different, the probe
Performance Verification
6
83