Keysight N2800A User Manual page 53

Keysight n2800a series infiniimax iii series probing system
Table of Contents

Advertisement

Replacing N5441A Probe Head Wires
Use the following procedure to install or replace the wire leads on the N5441A solder-in
probe head. Depending on your probing application, you can order either 5 mil or 7 mil
wire as listed in the following table. For example, use the 5 mil wire for attaching to small
vias.
Table 6
Wire Diameter
0.007 inch (tin-plated nickel wires)
0.005 inch (tin-plated nickel wires)
Table 7
Recommended Equipment
Equipment
Vise or clamp for holding tip
Metcal STTC-022 (600 °C) or STTC-122 (700 °C) tip soldering iron or equivalent. The 600 °C tip will help limit burning of the
FR4 tip PC board.
0.381 mm (0.015 in) diameter RMA flux standard tin/lead solder wire
Fine stainless steel tweezers
Rosin flux pencil, RMA type (Kester #186 or equivalent)
Flush cutting wire cutters
Magnifier or low power microscope
Keysight supplied trim gauge (01169-23801)
Procedure
1
Use the vise or clamp to position the tip an inch or so off the work surface for easy access.
If using a vise, grip the tip on the sides with light force. When tightening the vise, use light
CAUTION
force to avoid damaging the solder-in probe head If using a tweezers clamp, grip the tip either
on the sides or at the top and bottom.
InfiniiMax III Series Probes User's Guide
Required Wire Type
Maintaining Probe Heads
Part Number
01169-81301
01169-21306
3
53