6
Performance Verification
Figure 64 De-embedded "Vin" trace
10
Now move the 2.92 mm male test port 1 connector to the 5062_1247 adapter and N5477A
Autoprobe II adapter assembly. Connect the Maury Microwave 8775B2 2.92 mm male
broadband load to the 1250_1749 adaptor and N5443A PV fixture assembly. This new
setup is shown in
11
Install the proper file to de-embed the 5062_1247 adapter and N5477A adapter from port
1 of the VNA.
82
Figure
65. Torque all connections.
Create a Touchstone file for the 5062-1247 and N5477A by cutting and
a
pasting the text in
Name the file Adapter_5062_1247___Adapter_N5477A.s2p.
Go to menu "Calibration/Fixturing Selections/2 Port De-embedding" and
b
select Port 1.
Set S2P file selection to the file saved in step a.
c
Make sure the "Enable De-embedding" box is still checked.
d
Under "Calibration" menu, make sure "Fixturing ON/off" is still checked so
e
file is being used for de-embedding.
"Touchstone File (5062-1247 & N5477A)"
on page 92.
InfiniiMax III Series Probes User's Guide