Tsens_Type - Omega cl310 User Manual

M-4590/1107
Table of Contents

Advertisement

TSENS_TYPE?

Returns the type of sensor that is currently set to measure
temperature, either TC or RTD.
UPPER_MEAS
Sets the measuring mode for the upper display. After the command
enter DCI for mA, DCI_LOOP for mA with loop power, DCV for volts,
and PRESSURE for pressure. For example:
UPPER_MEAS DCV sets the upper display to measure volts
U_PRES_UNIT
Sets the unit for measuring pressure on the upper display. Add the
unit after the command. The available pressure units and their syntax
are shown in Table 7. (Parameter Units). For example:
U_PRES_UNIT MMHG sets the pressure unit to millimeters of
mercury at 0°C
VAL?
Returns the value of any measurement taking place on the upper
and lower display. For example, if the upper display is measuring
5mA, and the lower display is measuring 10V, then VAL? will return:
5.000000E-03, A, 1.000000E+01, V
ZERO_MEAS
Zeroes the attached pressure module. Enter the zeroing value in PSI
after the command when zeroing an absolute pressure module.
ZERO_MEAS?
Returns the zero offset or the reference value for absolute pressure
modules.
50