saving any data, press either the HOME key or the ABORT function
key. A prompt is displayed requesting confirmation that the test data
accumulated so far is to be abandoned. Press the NO function key
to return to the data collection process at the point where it was
interrupted. Press the YES function key to abandon the data
collected.
Pulse Train Type Prompt
When the input/output combination includes a pulse train output, the
following is displayed for each test point.
The current test type, AF for As Found, and point number are shown
in the upper right corner of the display. Note that the point number is
the next point to be tested and data has not been saved for it yet.
Enter the output value to be generated using the numeric keys,
followed by ENTER. Press the TRIG function key to generate the
pulse train. The lower line changes to the following while the pulse
train is being generated.
Press the STOP function key to abort the pulse train before it is
finished. When the pulse train is complete, or is aborted, the lower
line changes back to its original state.
Once the calibrator input value has settled, and the results are
acceptable, press the READY function key to proceed to the prompts
for saving the data.
If the calibrator input is manual, the lower line changes to the
following prompts.
Press the NEXT function key to enter the manual input value as
described in the subsection 'Manual Data Entry Prompt' below. See
69