SIM

Sets the output for current simulation. This command also switches
the calibrator into mA simulation mode. A number and a unit must be
entered after the command. For example:
SIM 5 MA sets the current simulation at 5 mA
SIM?
Returns the output of the current simulation. With the example
above, the output would be:
5.000000E-03, A
TAG_CLEAR tag
Clears test data for the specified tag from the calibrator. Once the
test data is cleared, it can not be recovered. Returns <Complete> if
successful, otherwise an error message enclosed in angle brackets.
For example:
TAG_CLEAR FT-8567

TAG_CLEAR_ALL

Clears test data for all tags from the calibrator. Once the test data is
cleared, it can not be recovered. Returns <Complete> if successful,
otherwise an error message enclosed in angle brackets.

TAG_DNLD

See the separate Tag Download and Upload Procedures manual for
detailed instructions on using this command to download tag
calibration test configuration data from a PC to the calibrator.

TAG_UPLD

See the separate Tag Download and Upload Procedures manual for
detailed instructions on using this command to upload tag calibration
test results to a PC from the calibrator.

TAGS?

Sends a list of all tag ID's in the calibrator to a PC. Returns a list of
tag ID's, one per line, followed by <Complete> if successful,
otherwise an error message enclosed in angle brackets. Each line
contains three data items separated from each other by spaces.
First, the position of the tag in the calibrator, an integer from 1 to 50.