Tsens_Type; Upper_Meas; U_Pres_Unit; Val - Omega PCL1200 User Manual

Portable, high accuracy multifunction calibrator
Table of Contents

Advertisement

TSENS_TYPE

Sets the temperature sensor type to thermocouple, or to RTD for
temperature measurement. After the command add TC for
thermocouple, or RTD for RTDs. For example:
TSENS_TYPE TC sets the sensor type to thermocouple
TSENS_TYPE?
Returns the type of sensor that is currently set to measure
temperature, either TC or RTD.

UPPER_MEAS

Sets the measuring mode for the upper display. After the command
enter DCI for mA, DCI_LOOP for mA with loop power, DCV for volts,
and PRESSURE for pressure. For example:
UPPER_MEAS DCV sets the upper display to measure volts

U_PRES_UNIT

Sets the unit for measuring pressure on the upper display. Add the
unit after the command. The available pressure units and their syntax
are shown in Table 7. (Parameter Units). For example:
U_PRES_UNIT MMHG sets the pressure unit to millimeters of
mercury at 0°C

VAL?

Returns the value of any measurement taking place on the upper
and lower display. For example, if the upper display is measuring
5mA, and the lower display is measuring 10V, then VAL? will return:
5.000000E-03, A, 1.000000E+01, V

ZERO_MEAS

Zeroes the attached pressure module. Enter the zeroing value in PSI
after the command when zeroing an absolute pressure module.
ZERO_MEAS?
Returns the zero offset or the reference value for absolute
pressure modules.
49