BB703
Blackbody Calibrator

Advertisement

Table of Contents
loading

Summary of Contents for Omega BB703

  • Page 1 User’s Gui d e Extended Warranty Program Shop online at omega.com e-mail: info@omega.com For latest product manuals: www.omegamanual.info BB703 Blackbody Calibrator...
  • Page 2 Fax: (203) 359-7700 e-mail: info@omega.com For Other Locations Visit omega.com/worldwide The information contained in this document is believed to be correct, but OMEGA accepts no liability for any errors it contains, and reserves the right to alter specifications without notice.
  • Page 3: Table Of Contents

    4.2.2 Target Plate ..................4-1 4.2.3 Fan ..................4-1 Fuse Replacement ................4-1 Section 5 Specifications ................5-1 Section 6 Troubleshooting Guide ..............6-1 Section 7 Glossary of Terms Used in This Manual ......... 7-1 The OMEGA Family of Blackbody Calibrators ....... 7-2...
  • Page 4 Table of BB703 Blackbody Calibrator Figures Table of Figures Figure Description Page IEC Symbols ................. 1-2 Limitation of Maximum Setpoint Temperature at Elevated Ambient Temperatures .............. 2-2 Front Panel ................... 3-1 Back Panel ..................3-2 Overheat Reset Switch ............... 3-3 Menu Hierarchy Showing Factory Default Settings ....
  • Page 5: Section 1 - Introduction

    BB703 Introduction Section 1 - Introduction Your BB700 Series Blackbody Calibration Source has been designed for ease of use and reliability whenever you have the need to test or calibrate non-contact infrared temperature instruments. It is important that you read this manual completely and follow all safety precautions before operating this instrument.
  • Page 6: General Description

    Figure 1. IEC symbols 1.3 General Description The Model BB703 is a portable, rugged, bench-top, hot blackbody calibration source with a built-in precision PID digital controller. The calibrator is used to test and calibrate infrared pyrometers. The 28.6mm (1.125") diameter target plate has an emissivity of .95 and can be set to any temperature between ambient +11...
  • Page 7: Mounting

    0 to 40°C (32 to 104°F). 2.3 Ambient Temperature The target plate of the BB703 can achieve any temperature within the specified temperature range of ambient +11 to 400°C (ambient +20 to 752°F) when being operated in ambient temperatures up to 25°C (77°F). When operating the unit at higher ambient temperatures, the user must not exceed the “Maximum...
  • Page 8: Power Connection

    2.4 Power Connection Standard (115 Vac~, 50/60 Hz models) The BB703 comes with a standard North American 3-prong AC power cord. Do not use any other power cord other than the one provided. This cord provides the proper grounding and has been safety tested by the proper safety agencies.
  • Page 9: Section 3 - Operation

    Keep flammable products such as paper, plastics and clothing far from the BB703. • The BB703 is a Class II instrument. It is intended to be operated in laboratory environment only. Never operate the unit outside or around children.
  • Page 10: Back Panel Connections

    NOTE: P.I.D. Control Proportional, integral, derivative control ( P.I.D.) is a temperature control algorithm used in high end temperature controllers. The controller causes the process to attain the desired temperature by turning the process on or off. The process may be a heater or refrigerator. As the process temperature approaches the setpoint temperature the hot or cold process will be pulsed to reduce the corrective measures and minimize overshooting.
  • Page 11: Overheat Reset Switch

    Figure 5. Overheat Reset Switch The layout of the front panel is shown in Figure 3. The BB703 incorporates a PID digital setpoint controller. In the default mode the digital display indicates the blackbody target plate temperature known as (PV) Process Variable. Holding down the modify key causes the display to show the programmed setpoint known as (SV) Setpoint Variable.
  • Page 12: Target Plate Temperature Transition Graph

    Figure 7. Programming Procedures 3.6 Approximate Target Plate Temperature Transition Graph The BB703 heat-up and cool-down characteristics are illustrated in the following graph. The graph illustrates the approximate time required to make a transition from one plate temperature to another. First determine the current plate temperature and go to the corresponding row.
  • Page 13: Maintenance

    2 ea. 250 Vac~, F1A (Fast-Acting, 1A) VDE APPROVED (5mm dia. x 20mm long). NOTE: Note that for the BB703-230 Vac both fuses need to be checked if there is a suspicion that a fuse has been blown. One or both fuses may have been damaged.
  • Page 14: Specifications

    BB703 Specifications Section 5 - Specifications Target Plate Temperature Range: Ambient + 10°C to 400°C* (Ambient + 20°F to 752°F)* Ambient Environmental Conditions Temperature: 0 to 40°C (32 to 104°F)* Humidity: 0 to 90% RH, non-condensing Power BB703: 115 Vac~, 50/60 Hz, 175W...
  • Page 15: Troubleshooting Guide

    BB703 Troubleshooting Guide Section 6 - Troubleshooting Guide Problem Solution 1. Unit will not turn on. a. Check all electrical connections. b. Check rear panel fuse(s). c. Unit requires service, contact our customer service department. 2. Unit turns on but the target a.
  • Page 16: Section 7 - Glossary Of Terms Used In This Manual

    BB703 Glossary of Terms Used in This Manual Section 7 - Glossary of Terms Used in This Manual Blackbody A theoretical object that radiates the maximum amount of energy at a given tem- perature, and absorbs all the energy incident upon it.
  • Page 17: The Omega Family Of Blackbody Calibrators

    The OMEGA Family of Blackbody Calibrators The OMEGA Family of Blackbody Calibrators Listed below is a selection guide of OMEGA’s current line of blackbody calibration sourc- es in addition to the one you have selected. This family of rugged, portable and accurate calibrators cover a wide range of temperatures, target plate sizes and features making them perfect for infrared pyrometer field service testing and laboratory calibrations.
  • Page 18 Notes:...
  • Page 19 Department will issue an Authorized Return (AR) number immediately upon phone or written request. Upon examination by OMEGA, if the unit is found to be defective, it will be repaired or replaced at no charge. OMEGA’s WARRANTY does not apply to defects resulting from any action of the purchaser, including but not limited to mishandling, improper interfacing, operation outside of design limits, improper repair, or unauthorized modification.
  • Page 20 Where Do I Find Everything I Need for Process Measurement and Control? OMEGA…Of Course! Shop online at omega.com TEMPERATURE M U Thermocouple, RTD & Thermistor Probes, Connectors, Panels & Assemblies M U Wire: Thermocouple, RTD & Thermistor M U Calibrators & Ice Point References M U Recorders, Controllers &...