4.7 Sourcing a pulse train
The calibrator can produce a pulse train with an adjustable number
of pulses at a desired frequency. For example, setting the frequency
to 60Hz and the number of pulses to 60 would produce 60 pulses for
a period of 1 second. To source a pulse, use the same connection as
for frequency, and proceed as follows:
1. Switch to the lower display and select pulse from the primary
parameters.
2. Choose the desired unit and enter the frequency using the
keypad.
3. Select the [COUNTS] function from the home menu to enter the
number of pulses. Use [TRIG] to start and stop the signal.
4. The amplitude of the pulse can be adjusted in the same manner
as for frequency.
Figure 13. Connections for Outputting Thermocouples
4.8 Sourcing Thermocouples
To source a thermocouple use the following steps:
1. Connect the thermocouple leads to the appropriate polarized TC
miniplug, and insert the plug into the TC terminals on the
calibrator, as shown in Figure 13.
2. Select lower display from the Main Menu, and choose
thermocouple [TC] from the primary parameters.
3. Choose output [OUT] from the input/output control.