Press the TEST function key to display the test selection menu.
Press the AS LEFT function key to display the tag selection prompt
which shows all previously calibrated tags available for recalibration.
Tag numbers are displayed on multiple pages, six tags to a page.
Press the UP ARROW and DOWN ARROW keys to scroll the tag
selector highlight one tag at a time, scrolling from page to page,
cycling from top to bottom of the list and vice versa. Use the LEFT
ARROW and RIGHT ARROW keys to scroll the tag selector highlight
one page at a time, cycling from top to bottom of the list and vice
versa.
Press the EXIT function key to return to the main menu without
selecting a tag.
When the desired tag number is highlighted, press the SELECT
function key.
If the there has been no previous As Left test proceed directly to data
collection.
If there has been at least one previous As Left test, a prompt
requesting confirmation to overwrite existing As Left data is
displayed. Press the YES function key to continue to data collection.
Press the NO function key to return to tag selection.
8.5.2 Data Collection
If pressure is selected on the upper or lower display, the pressure
zero prompt is displayed first, and operates as described under As
Found Setup.
As Left data collection is identical to As Found data collection except
for four minor differences.