Event Status Register (ESR)
A two-byte register, in which the lower bits represent conditions of
the Calibrator. Cleared when read and when power is reset.
Event Status Enable Register (ESE)
Enables and disables bits in the ESR. Setting a bit to 1 enables the
corresponding bit in the ESR, and setting it to 0 disables the
corresponding bit. Cleared at power reset. Bit assignments for the
ESR and the ESE respectively are shown below.
15
0
7
PON
PON
CME
EXE
DDE
QYE
34
14
13
0
0
6
5
0
CME
EXE
Power On. Set to 1 if power was turned on and off before
the Event Status Register was read.
Command Error. Set to 1 when the calibrator receives an
invalid command. Entering an unsupported RTD type may
cause such an error.
Execution Error. Set to 1 when the calibrator runs into an
error while executing is last command. A parameter that
has too significant figures may cause such an error.
Device-dependent Error. Set to 1 when, for example, the
output of the calibrator is overloaded.
Query Error.
12
11
0
0
4
3
DDE
QYE
10
9
0
0
2
1
0
OPC
8
0
0