REMOTE

Puts the calibrator in remote mode. From the remote mode the user
can still use the keypad to get back to local unless the command
LOCKOUT was entered before REMOTE. Than the keypad is totally
locked out, and the user has to send the LOCAL command to get
back to local operation.

RTD_TYPE

Sets the RTD type. The following is a list of RTD types they way they
should be entered after the command:
PT385_10;
PT385_500;
Ni120;
OHMS;
For example:
RTD_TYPE PT385_10 sets RTD type to Pt385-10
RTD_TYPE?
Returns the RTD type.

RTD_WIRE

Sets the number of wires used for connection in measuring RTDs.
The calibrator measures RTDs using 2-, 3-, and 4-wire connections.
After the command, enter 2W for 2- wire, 3W for 3-wire, and 4W for 4-
wire. For example:
RTD_WIRE 4W sets the connection to 4-wire
RTD_WIRE?
Returns the number of wires used in the RTD connection.
SET_CLOCK yyyy mm dd hh nn ss
Sets the calibrator date and time, where yyyy is the four digit year
(2006 to 2100), mm is the two digit month (1 to 12), dd is the two
digit day (1 to the number of days in the specified month), hh is the
two digit hour in 24 hour format (0 to 23), nn is the two digit minute
(0 to 59), and ss is the two digit second (0 to 59). Note that all two
digit values must have a leading zero if they have a value less than
10. Returns <Complete> if successful, otherwise an error message
enclosed in angle brackets. For example:

SET_CLOCK 2006 03 20 09 16 33

PT385_50;
PT385_1000;
Cu10;
Cu50;
CUSTOM;
PT385_100;
PT392_100;
Cu100; YSI_400;
PT385_200;
PTJIS_100;