*ESE?
Returns the contents of the Event Status Enable register. The value
returned is a decimal. For example, if the register has the following
settings:
10000101
*ESR?
Returns the contents of the Event Status Register in decimal form.
For example:
If the ESR contains 10111001, *ESR? will return 185.
*IDN?
Returns the manufacturer, model number, and firmware revision of
the Calibrator. For example:
*IDN? will return OMEGA,CL310,0,2.00
*OPC
Enables the Operation Complete setting in the ESR. This setting
makes it possible to check if an operations is complete after it has
been initialized.
For example this operation could be used with the command TRIG.
*OPC?
Returns 1 when all operations are complete, and causes program
execution to pause until all the operations are complete. For
example:
TRIG ; *OPC? will return a 1 when the pulse train initiated by
TRIG is complete.
*RST
Resets the state of calibrator to the power-up state. All subsequent
commands are held off until the execution of the command is
complete.
*SRE
Loads a byte into the Service Request Enable register. A decimal
number must be entered, which when converted to binary,
corresponds to the correct settings.
For example:
*SRE 8 enters the binary number 00001000 to the SRE. This
40
enables bit 3. Bit 6 is not used.
than the value returned will be 133.