These two menus operate the same as the normal upper and lower
type and parameter selection menus described earlier in this manual,
except that they have two extra type selections, MANUAL IN and
MANUAL OUT. The currently selected parameter field blinks
(indicated by bold italicized text above). Press the SELECT function
key to cycle through the possible selections for that field. Press the
NEXT function key to move to the next parameter field. When all
parameters are set for the selected display, press the DONE function
key to return to the As Found setup menu.
One display must be selected as a calibrator input type, and the
other must be selected as a calibrator output type. Pressure can not
be selected simultaneously on both displays as the calibrator is only
capable of handling one pressure at a time. Pressure takes its input
or output orientation as the opposite of the orientation selected on
the other display.
Once both the upper and lower displays are properly selected, press
the DONE function key on the As Found setup menu to proceed with
the setup.
At this point some display type selections, as described below,
require optional setup data to be entered. Otherwise pressing DONE
proceeds directly to the As Found data collection prompt.
Optional Setup - Manual Input or Output on Upper Display
If either manual type is selected on the upper display, the upper
manual unit description prompt is displayed.
64