Omega CL1000 Series Quick Start Quide

Omega CL1000 Series Quick Start Quide

Mono hot point dry block probe calibrator
Hide thumbs Also See for CL1000 Series:

Advertisement

Quick Links

WARRANTY/DISCLAIMER
OMEGA ENGINEErING, INC. warrants this unit to be free of defects
in materials and workmanship for a period of 13 months from
date of purchase. OMEGA's WArrANTY adds an additional one (1)
month grace period to the normal one (1) year product warranty
to cover handling and shipping time. This ensures that OMEGA's
customers receive maximum coverage on each product.
If the unit malfunctions, it must be returned to the factory for
evaluation. OMEGA's Customer Service Department will issue an
Authorized return (Ar) number immediately upon phone or written
request. Upon examination by OMEGA, if the unit is found to be
defective, it will be repaired or replaced at no charge. OMEGA's
WArrANTY does not apply to defects resulting from any action of
the purchaser, including but not limited to mishandling, improper
interfacing, operation outside of design limits, improper repair, or
unauthorized modification. This WArrANTY is VOID if the unit
shows evidence of having been tampered with or shows evidence of
having been damaged as a result of excessive corrosion; or current,
heat, moisture or vibration; improper specification; misapplication;
misuse or other operating conditions outside of OMEGA's control.
Components in which wear is not warranted, include but are not
limited to contact points, fuses, and triacs.
OMEGA is pleased to offer suggestions on the use of its various
products. However, OMEGA neither assumes responsibility for
any omissions or errors nor assumes liability for any damages
that result from the use if its products in accordance with
information provided by OMEGA, either verbal or written.
OMEGA warrants only that the parts manufactured by the
company will be as specified and free of defects. OMEGA
MAKES NO OTHER WARRANTIES OR REPRESENTATIONS OF
ANY KIND WHATSOEVER, EXPRESSED OR IMPLIED, EXCEPT
THAT OF TITLE, AND ALL IMPLIED WARRANTIES INCLUDING
ANY WARRANTY OF MERCHANTABILITY AND FITNESS
FOR A PARTICULAR PURPOSE ARE HEREBY DISCLAIMED.
LIMITATION OF LIABILITY: The remedies of purchaser set forth
herein are exclusive, and the total liability of OMEGA with
respect to this order, whether based on contract, warranty,
negligence, indemnification, strict liability or otherwise,
shall not exceed the purchase price of the component upon
which liability is based. In no event shall OMEGA be liable for
consequential, incidental or special damages.
CONDITIONS: Equipment sold by OMEGA is not intended to be used,
nor shall it be used: (1) as a "Basic Component" under 10 CFr 21 (NrC),
used in or with any nuclear installation or activity; or (2) in medical
applications or used on humans. Should any Product(s) be used in or
with any nuclear installation or activity, medical application, used on
humans, or misused in any way, OMEGA assumes no responsibility
as set forth in our basic WArrANTY/DISCLAIMEr language, and,
additionally, purchaser will indemnify OMEGA and hold OMEGA
harmless from any liability or damage whatsoever arising out of the use
of the Product(s) in such a manner.
RETURN REQUESTS/INQUIRIES
Direct all warranty and repair requests/inquiries to the OMEGA
Customer Service Department.
BEFOrE rETUrNING ANY
PrODUCT(S) TO OMEGA, PUrCHASEr MUST OBTAIN AN
AUTHOrIZED rETUrN (Ar) NUMBEr FrOM OMEGA'S CUSTOMEr
SErVICE DEPArTMENT (IN OrDEr TO AVOID PrOCESSING
DELAYS). The assigned Ar number should then be marked on the
outside of the return package and on any correspondence.
FOr WARRANTY rETUrNS,
FOr NON-WARRANTY rEPAIrS,
consult OMEGA for current repair
please have the following
charges. Have the following
information available BEFOrE
information available BEFOrE
contacting OMEGA:
contacting OMEGA:
1. Purchase Order number under
1. Purchase Order number to
which the product was
cover the COST of the repair or
PUrCHASED,
calibration,
2. Model and serial number of
2. Model and serial number of the
the product under warranty, and
product, and
3. repair instructions and/or
3. repair instructions and/or specific
specific problems relative to
problems relative to the product.
the product.
OMEGA's policy is to make running changes, not model changes,
whenever an improvement is possible. This affords our customers
the latest in technology and engineering.
OMEGA is a registered trademark of OMEGA ENGINEErING, INC.
© Copyright 2017 OMEGA ENGINEErING, INC. All rights reserved.
This document may not be copied, photocopied, reproduced,
translated, or reduced to any electronic medium or machine-readable
form, in whole or in part, without the prior written consent of OMEGA
ENGINEErING, INC.
MQS3529/0417
5
with the
(up arrow) button or
(Down arrow)
button. Holding the
button and the
(Up)/
(Down) button for an extended period will cause
the setpoint temperature to advance more rapidly to a
desired value. The minimum and maximum setpoints
are locked out at 0 and 260°C (32 and 500°F),
respectively.
Changing the Display Units from °C to °F or
(°F to °C)
Press and hold the
(Up) and
(Down) buttons on
the controller at the same time for 3-4 seconds, then
release. The display will be flashing between "tune –
off". Press the
button one time. The display will be
flashing between "LEVEL-1". While holding the
(Modify) key down press the
button one time. The
display will be flashing between "LEVEL-2". Now that
you are on menu level 2 release the
button. Press
the
button eleven times. The display will now be
flashing between "unit-°F" or "unit-°C". To change the
current units setting press and hold the
button,
then press the
or
button one time to change
between units. To save your change and return to the
run mode press and hold the
and
buttons at the
same time for 3-4 seconds, then release.
Communication (RS-232)
This section only applies to CL1000 Series Calibrators
purchased with optional RS-232 Communications
(-C2) Option
Your calibrator unit has been factory pre-wired
and configured for ease of use with a stereo jack
connection on the side panel that will require no
additional wiring. An interface cable, Part No.
OM-NOMAD-CP9 is included with your unit for easy
connection between your benchtop calibrator and PC.
BENCHTOP CONTROLLER
STEREO JACK SIDE PANEL
INTERNAL WIRING
TX
RX
STEREO PLUG –– DB–9
GND
INTERFACE CABLE
TX
1
RX
GRN
1.8 m (6 ft.)
CONTROLLER
PC COM PORT
10
11
13
13
14
15
16
2
3
4
5
6
7
8
Wiring Connections From Calibrator To PC
ARTWORK/PRODUCT ART/
DWGS/START HERE ARROW
6
Calibrators with the –C2 option come complete with
communication software, OMEGA part number CN9-
SW. The software is designed to interface with your
calibrator.
NOTE:
The CN9-SW software is Windows 95/98/NT/2K/
XP compatible.
Cool-Down Procedure
1. Turn the power switch on the back of the calibrator
to "standby" mode.
2. The cooling fan may continue to run or turn on
after several minutes if the heater assembly inside
the unit has not yet fully cooled down to ambient
temperature. Do not remove the power cord or
the main line power until the fan stops running
permanently.
3. It is now safe to remove your unit from the power
source and/or return to storage.
Overheat Reset Switch
If the unit is operated at high temperatures in elevated
ambient temperatures, an overheat condition may
occur. In an overheat situation a mechanical reset
switch on the right side panel will pop and open the
heater circuit. The controller will still have power.
While the controller will be demanding heat from the
heater, the process temperature will fall continuously
until it equalizes with room temperature. If an
overheat condition occurs, let the unit cool off for one
hour (leave the unit on), then press the reset button,
firmly. If this does not correct the problem, contact the
factory.
OVERHEAT RESET SWITCH
FRONT
HANDLE (IN RETRACT POSITION)
5
Overheat Reset Switch (Right Side Panel)
DB–9
For complete product manual:
www.omega.com/manuals/manualpdf/M3529.pdf
CL1000 SERIES
Mini hot point
®
Block Probe Calibrator
omega.com info@omega.com
Servicing North America:
U.S.A.:
Toll-Free: 1-800-826-6342 (USA & Canada Only)
Customer Service: 1-800-622-2378 (USA & Canada Only)
Engineering Service: 1-800-872-9436 (USA & Canada Only)
Tel: (203) 359-1660
e-mail: info@omega.com
Canada:
Toll-Free: 1-800-826-6342 (USA & Canada Only)
Tel: (514) 856-6928
e-mail: generalinfo@omega.ca
Servicing Mexico and Latin America:
Mexico/
Tel: 001-800-099-0420 (Mexico Only)
Latin America
(Outside Mexico) Fax: 001 (203) 968-7290
e-mail: ventas@mx.omega.com
Servicing Asia:
China:
Hotline: (+86) 800 819 0559, (+86) 400 619 0559
e-mail: info@cn.omega.com
Servicing Europe:
FAN
France:
Freephone: 0805 541 038 (France only)
Tel: 01 57 32 48 17
e-mail: esales@omega.fr
Germany/
Freephone: 0800 826 6342 (Germany only)
Austria:
Tel:+ 49 (0)7056 9398-0
BACK
e-mail: info@omega.de
Italy
Freephone: 800 906 907 (Italy only)
Tel: +39 022 333 1521
e-mail: commerciale@it.omega.com
Netherlands:
Freephone: 0800 099 33 44 (Netherlands only)
Benelux
Tel: +31 070 770 3815
e-mail: esales@omega.nl
Spain:
Freephone: 800 900 532 (Spain only)
Tel: +34 911 776 121
e-mail: ventas@es.omega.com
United
Freephone: 0800 488 488 (United Kingdom only)
Kingdom:
Tel: +44 (0)161 777 6611
e-mail: sales@omega.co.uk
It is the policy of OMEGA Engineering, Inc. to comply with all worldwide safety
and EMC/EMI regulations that apply. OMEGA is constantly pursuing certification of
its products to the European New Approach Directives. OMEGA will add the CE
mark to every appropriate device upon certification.
The information contained in this document is believed to be correct, but
OMEGA accepts no liability for any errors it contains, and reserves the right to
alter specifications without notice.
WARNING: These products are not designed for use in, and should not be used
for, human applications.
Dry
®
Fax: (203) 359-7700
Fax: (514) 856-6886
Web: www.omega.com
Tel: 001-203-359-7577
Web: mx.omega.com
Web: cn.omega.com
Fax: 01 57 32 48 18
Web: www.omega.fr
Fax: +49 (0) 7056 9398-29
Web: www.omega.de
Fax: +39 022 333 1522
Web: it.omega.com
FAX: +31 070 770 3816
Web: www.omega.nl
Fax: +34 911 776 122
Web: es.omega.com
Fax: +44 (0)161 777 6622
Web: www.omega.co.uk

Advertisement

loading

Summary of Contents for Omega CL1000 Series

  • Page 1 Wiring Connections From Calibrator To PC OMEGA is a registered trademark of OMEGA ENGINEErING, INC. It is the policy of OMEGA Engineering, Inc. to comply with all worldwide safety © Copyright 2017 OMEGA ENGINEErING, INC. All rights reserved. and EMC/EMI regulations that apply. OMEGA is constantly pursuing certification of This document may not be copied, photocopied, reproduced, its products to the European New Approach Directives.
  • Page 2 ARTWORK/PRODUCT ART/ ARTWORK/PRODUCT ART/ ARTWORK/PRODUCT ART/ ARTWORK/PRODUCT ART/ DWGS/START HERE ARROW DWGS/START HERE ARROW DWGS/START HERE ARROW DWGS/START HERE ARROW Start Here NOTE: Using This Quick Start Manual Power Connection Probe Well Styles • Hole in center of probe well is a vent hole, not Connect the unit to a stable, earth grounded AC Use this Quick Start Manual with your CL1000 a probe hole.