Omega cl310 User Manual page 81

M-4590/1107
Table of Contents

Advertisement

The PT in the upper right corner of the display indicates that this is
the adjustment display. If an As Found data point is selected, its
number is displayed. If a user entered value is selected, two dashes
are displayed.
Use the numeric keys, followed by ENTER, to enter a user value for
generation, or press the READY function key to proceed to the next
menu where the UP and DOWN function keys can be used to scroll
through the As Found data points.
Press the TRIG function key to generate the pulse train. The lower
line changes to the following while the pulse train is being generated.
Press the STOP function key to abort the pulse train before it is
finished. When the pulse train is complete, or is aborted, the lower
line changes back to its original state.
Press the READY function key to change the lower line to the
prompts for scrolling through the As Found test points, or
proceeding to As Left data collection.
Press the UP or DOWN function key to cycle through the As Found
test points.
Press the AS LEFT function key to proceed directly to As Left data
collection.
Press the LEFT ARROW key to return to the trigger menu and
generate a new pulse train.
77