1. Select a unit to validate. (Fixed parameters are used as titration parameters.)
CA-200
CA-200/VA-200
CA-200/VA-210
CA-200/VA-124S
CA-200/VA-121
CA-200/VA-122
CA-200/VA-230
2. Input a theoretical value by μg or % unit.
Example: 5000μg Theoretical value: [5] key + [0] key + [0] key + [0] key + [Enter] key
3. Input tolerance by μg or % unit.
Example: 100μg Tolerance: [1] key + [0] key + [0] key + [Enter] key
4. After setting, press [Escape] key to turn to the validation titration window.
5. The contents set at (1) to (4) are printed.
POINT
Recommended values are included in titration parameters and they can be changed if
necessary. Change the formula "Calc Form" to the formula for standard sample.
Unit are set by selected formulas.
6. Measure moisture. Refer to 5. Moisture Measurement Mode.
7. After measurement, result is indicated and printed.
To repeat measurement, select "Yes" to return to the titration window.
Select "No" to print statistical calculation result and the judgement.
(Only when all results are within tolerance ranges, "OK" is printed.)
POINT
If normal measured values can't be obtained by these methods, electrolyte or detection
electrodes or working electrodes or moisture meters or parameters can be wrong.
Operators can't calibrate the unit, therefore contact our distributors.
Refer to Section 9: Troubleshooting.
Measurement validation
01 System
02 Theo conc
02
03 Tolerance
Validation system
6-23
Section 6: Function
: CA-200
: 5000
: 100
** CHANNEL 1 **
<Moisture Validation>
1.System: KF- 200
Data: 2006/05/23
Time : 15:33
2.Moisture Check
Theo Conc: 0
Tolerance: 0