Recalculation; Statistical Calculation - Mitsubishi CA-200 Instruction Manual

Moisture meter
Table of Contents

Advertisement

CA-200

6-1-2. Recalculation

The following changes are as follows at recalculation.
1. Print format, print cycle change (possible only for the last data)
By changing a print cycle when format is 5, printing by the changed print cycle is possible.
2. Concentration calculation formula and unit change
Based on measured moisture content (μg), results calculated by changed formulas and units are
printed.
3. Sample parameter change
Results calculated by changed values are printed.
(1) Move to items to recalculate with [↑] and [↓] keys and press [Enter] key.
* Also selectable by inputting ID No. directly with numerical keys and pressing [Enter] key.
(2) Move to items to change with [↑] and [↓] keys.
(3) Change "Print Form" or "Calc Form" with [←] and [→] keys.
Input "W" and "w" such as sample parameters with numerical keys.
(4) Press [Enter] key to determine the contents.
(5) Press [Print] key to print them. The changed contents are saved.
* By pressing [Escape] key without pressing [Print] key, the changed contents are not saved.
The window returns to the result list.

6-2. Statistical calculation

Statistical calculations are internal statistical calculation and external calculation.
Internal statistical calculation: Statistical calculation of specified data in the last 200 data
External statistical calculation: Statistical calculation of manually-inputted data independent of
saved data
6-3
Section 6: Function