5-6-2. Recovery measurement

To check that the measurement whole including the unit and reagents is normal, measure a standard
sample by the same method as sample measurement. Use water or commercial water standard solution
as a standard sample.

5-6-3. Sample measurement

This titration is usually applied.
(1) Wait until background is stable. By pressing [Titration] key, dehydration is run and background is
down.
(2) Wait until "Stable" is displayed.
(3) After pressing [Start/Stop] key, inject a sample. Background is saved.
(4) Titration starts.
(5) Set sample parameters such as sample names and sample weight. (You can set sample names before
sample injection.)
(6) When measurement reaches the end point at background +End Sense, a buzzer beeps and the result
is printed out.
(7) Subsequently, repeat (2) to (6).
POINT
Press [Start/Stop] key to end measurement forcibly.
"*" is added to results. When a background value is lower at sample injection, result is
stable.

5-7. Measurement end

1. Do the following items if necessary.
Statistical calculation (6-2. Statistical calculation)
Recalculation (6-1. Recalculation)
Arrangement around the unit
2. Turn off the power switch.