Section 6: Function
(3) Alarm Set
When reagent "measured moisture content" is more than the set value, the check window is displayed
and the alarm beeps. Set the alarm to "No" so that this function is invalid. (Default)
Set the alarm to "Yes" so that this function is valid. Set (4) and (5).
(4) Anode solution preset volume
Set "measured moisture content" of anode solution. (Input range: 0 to 9999mg)
AQUAMICRON AX : 800
AQUAMICRON AKX : 300
(5) Cathode solution preset volume
Set "measured moisture content" of cathode solution. (Input range: 0 to 999mg)
AQUAMICRON CXU : 150

6-4. File Setting

Settings are saved even after power-off.

6-4-1. File protect setting

When a file is protected, titration file contents can't
be changed.
In addition, a multi-file can't be transferred.
(1) Protection setting
By pressing [Enter] key at the unprotected file
position, "Protect" is indicated and a file is protected.
(2) Protection release
By pressing [Enter] key at the protected file position, "Protect" is canceled and file protection is
released.
File protect setting
01 A.B.S
02 Acet.acid :Protect
03 Acetone
04 Acetonitr :
Protect was selected by Enter key.
6-6
01- 20[01]
:Protect
CA-200