CA-200

Section 5: Moisture Measurement Mode

5-1. Moisture measurement mode flow

(1) Turn on the power switch.
(2) Condition electrodes and dry a titration
flask. Press [Titration] key to condition
electrodes. After conditioning, dry the titration
flask inside.
(3) Selection a file.
Without changing a file, proceed to (4).
Press [Parameter/Character] key to display
the file selection window.
Select a file with [↓] [↑] keys+ [Enter] key or
number keys + [Enter] key.
Press [Escape] key to return to the
titration window.
(4) Sampling
Refer to "5-5. Sampling".
(5) Input a sample name and a measurement
number. (5-4-3. Sample parameters setting)
(6) Input "W" (sample weight + tare weight).
Press [Sample] key, input with number keys,
and press [Enter] key.
Press [Escape] key to return to the
titration window.
* Balance transfer input is possible at the titration window.
Mass is indicated in the message window.
Section 5: Moisture Measurement Mode
Standby
Time
Potential
Moisture
File Select
01 A.B.S
02 Acet.acid
03 Acetone
04 Acetonitr
Select with ↑, ↓ and press Enter.
Sample Parameter
01 Name
02 No
03 W
04 w
Sample Name
5-1
F01-01 Sample A
B.G.
mV
01-20[01]
F03 01-04
: Sample A
: 01
: 0.01
: 0.0075
Press Character key.
μg/s
μg

Advertisement