9-3. Error messages

When errors occur, the following messages are displayed.
When errors are not cancelled even by the following measures, contact our local distributors.
Error contents: Cathode solution consumption error
Causes and measures:
Cathode solution consumption is over preset values.
Change cathode solution and reset consumption to zero.
Error contents: Anode solution consumption error
Causes and measures:
Anode solution consumption is over preset values.
Change anode solution and reset consumption to zero
Error contents: Detection electrode error
Causes and measures:
A detection electrode can be disconnected or removed.
Check that a detection electrode is soaked in reagent.
For disconnection, change it.
Error contents: Cathode solution error
Causes and measures:
This error occurs when detection potential is up and titration speed is not up.
A cathode solution cell can be disconnected or removed.
For disconnection, change the cell.
Error contents: Divided by Zero
Causes and measures:
A concentration calculation denominator is zero.
Input numbers except a zero or change a calculation formula.
9-8