Section 5: Moisture Measurement Mode
(3) Min. Titr: Titr continuing time
For continuing titration forcedly by the set time
When "Delay" is set, titration continuing time
starts after "Delay".
Example: 5 minutes
Min Titr: [5] key + [Enter] key
(4) Titr Stop: Titration forced end time
Titration is forcedly stopped at the set time.
When setting "Delay", titration forced end
time is from after Delay end.
When setting shorter time than Min Titr,
it is invalid.
Example: 30 minutes
Min Titr: [3] key + [0] key + [Enter] key
(5) End Sense: End point sense
For determining the end point
When electrolysis speed during titration is End
Sense + background at titration start, titration
reaches the end point.
Titration
speed
Sample
injection
Start
Integrated
moisture
amount
Min Titr
Delay
Titration start
Illustration 5-2. Titration parameters
Titration Parameter F03 01-14
01 File Name : Acetone
02 Delay
03 Min Titr
04 Titr Stop : 0
Titration continuing time
(0)min
Titration Parameter F03 01-14
01 File Name : Acetone
02 Delay
03 Min Titr
04 Titr Stop
Titration forced end time
Titration Parameter F03 01-14
02 Delay
03 Min Titr
04 Titr Stop : 30
05 End Sense : 0.1
End Sense
Detected potential (mV)
Titration speed (μg/sec)
Titr Stop
Titration end
5-8
: 2
: 5
: 2
: 5
: 30
0-99(0)min
: 2
: 5
0-1 (0.1) μg/s
End Sense
B.G.
Time