Section 5: Moisture Measurement Mode
(9) Calc Unit: Result unit
Select a result unit from the following list.
Example: 0_%, ppm Auto Select
Calc Unit: [0] key + [Enter] key
Indications
0: %, ppm
Auto Select
1: % only
2: ppm only
3: Not Print
4: Option Unit
5: Option Unit
The first figure (0-5) shows the unit number.
(10) Blank Test: The number of times of
blank measurement
(Displayed when the formula with B is
selected.)
Specify the number of times of blank
measurement.
When the specified number of times of
measurement is completed, the average of
blank measurement is automatically
printed out and inputted in "B".
Subsequently, sample measurement can be run.
When a blank value is known, input "0" as the number of times and input the blank value in "B" of
sample parameters.
Example: Once
Blank Test: [1] key + [Enter] key
(11) VA Select: VA selection
Set the use and the type of water vaporizers.
Refer to vaporizer instruction manuals for
details.
Less than 0.1%: ppm is printed.
0.1% or more: % is printed.
Only % is printed.
Only ppm is printed.
No unit print
1
Except the above units, an optional unit is prepared.
2
Table 5-4. Result units list
Titration Parameter F03 01-14
06 Print Form : 5_R+S+P+G+I
07 Print Intv : 1
08 Calc Form : 1_M/(W-w)
09 Calc Unit : 0_%, ppm Auto Select
Calculation Unit
Contents
Titration Parameter F03 01-14
07 Print Intv : 1
08 Calc Form : 1_M/(W-w)
09 Calc Unit : 0_%,ppm Auto Select
10 Blank Test : 1
←→ Times of blank measure
Titration Parameter F03 01-14
18 Calc Form : 1_M/(W-w)
19 Calc Unit : 0_%, ppm Auto Select
20 Blank Test : 1
21 VA Select : 1_VA-200
VA Selection
5-12
CA-200
0-5(0)
0-10(1)
0-11(1)