Section 2: Installation and Assembly

2-2. Assembly

2-2-1. Titration cell

Remove each part of a titration cell and spread lightly contacting parts with attached exclusive grease.
Put cotton balls and silica gel into a drying tube.
POINT
Necessarily use exclusive grease. Particularly when moisture adsorbs to the ceramic
septum of a cathode solution cell, much time is required until the widow condition
indication is stable. After leaving a cell in the air for a long time, dry it with a dryer and
assemble it.
(1) Put a titration cell to a stirrer stand so that a sample injection plug is at the front.
(2) Insert the BNC connector 1 of a detection electrode into the right connection jack (green). Insert
the BNC connector 2 of a cathode solution cell into the left connection jack (blue). Turn the cell
right to fix it.
2
Illustration 2-1. Titration flask assembly
2-6
1
CA-200