File Selection; File Composition - Mitsubishi CA-200 Instruction Manual

Moisture meter
Table of Contents

Advertisement

Section 5: Moisture Measurement Mode

5-3. File selection

5-3-1. File Composition

Select a file for sample measurement. When "Dehydrating" or "Ready" or "Stable"is displayed,
a file can be selected. When "Delay" or "Titration" is displayed, a file cannot be selected.
(Titration and sample parameters during measurement can be displayed and set.)
When a CH2 P.C.B is installed, CH1 and CH2 have respective files.
File
03
02
01
Titration
parameters
File Name
Delay
Calc Form
Calc Unit
Temp Intv
Sample
parameters
Name
No
W
w
Y
Files
: They include titration parameters and sample parameters.
Up to 20 files can be saved.
Multi-files
: 50 files of our recommended titration parameters are set.
Transfer them to a file to use.
Concentration : They include 10 fixed formulas and 2 option formulas.
calculation
Specify them in titration parameters.
formulas
Results units
: They include 4 fixed units and 2 option units. Specify them in titration
parameters.
20
Transfer to a file
Illustration 5-1. File composition
50
Multi-file
03
02
01
Titration
parameters
File Name
Delay
Calc Form
Calc Unit
Unit
0 : Automatic
conversion
1 : % fixed
2 : ppm fixed
3 : Blank
4 : Option unit
5 : Option unit
5-4
CA-200
Calculation
Formula
0:M
1:M/(W-w)
2:M*X/(W-w)
3:M/(L*S)
4:M*Y*(1+t/273)/G
19: Optional formula

Advertisement

Table of Contents
loading

Table of Contents