CA-200

5-4-3. Sample parameters setting

Set sample parameters. Press [Sample] key to display only required items selected from the calculation
formula of titration parameters. Even if titration parameters are protected, they can be changed. All
sample parameters can be set or changed before and during measurement.
Parameters except "W" and "w" can be used successively every sample.
Input "W" and "w" every sample. If they are not inputted by measurement end, input is waited.
POINT
Set parameters "L", "S", "G", "t", "B", "X" and "Y" necessarily before the
measurement end. Once parameters are set, they are used successively every samples.
Sample parameters
Indications
Name:
No:
W:
w:
L:
S:
G:
t:
B:
X:
Y:
Function names
Sample name
Measurement No. (Automatic)
Sample weight + tare weight
wweight
Tare weight
Sample volume
Sample specific gravity
Gaseous sample volume
Gas temperature
Blank value
Optional coefficient X
Optional coefficient Y
Table 5-5. Sample parameters
Section 5: Moisture Measurement Mode
Max.
Characters 13 Characters 0
99
999
999
99
9
99
199
999
10000
10000
5-13
Min.
Units
1
0
g
0
g
0
ml
0
0
L
0
C
μg
0
0
0