Reagent Consumption Setting - Mitsubishi CA-200 Instruction Manual

Moisture meter
Table of Contents

Advertisement

CA-200
Example of statistical calculation
of internal data
* Internal statistical *
ID 07
ID 08
ID 09
ID 10
N =
X =
S =
RSD=
POINT
The data number of statistical calculation is from 2 to 200.

6-3. Reagent consumption setting

Reagent change is indicated in the window or by
the alarm. (At dehydration start and at
measurement end)
The setting is as follows.
(1) Anode solution consumption
Anode solution consumption is added automatically including dehydration and titration of measured
moisture content. (mg) At reagent change, set anode solution consumption to "0" to reset it
(2) Cathode solution consumption
Cathode solution consumption is added automatically including dehydration and titration of measured
moisture content. (mg) At reagent change, set cathode solution consumption to "0" to reset it
0.5002 %
0.4989 %
0.4970 %
0.4952 %
4
0.4978
0.0022
0.43 %
6-5
Example of statistical calculation
of external data
* External Statistical*
EX 1
EX 2
EX 3
EX 4
EX 5
N =
X =
S =
RSD=
Reagent consumption
001 Anode consu
002 Cath consu
003 Alarm set
004 Anode alarm
Reset by 0 key.
Section 6: Function
582
576
580
601
590
5
585.8
9.9096
1.69 %
: 0
: 0
: 0: No
: