9

Data output RS 232 C

9.1 Technical data

7 / 8 data bits, 1 / 2 stop bit, no / even / odd parity bit
Baud rate selectable at 2400, 4800, 9600, 19200 and 38400 Baud
Miniature plug-in necessary (9 pole D-Sub)
For operation with interface faultless operation is only ensured with the correct
KERN – interface cable (max. 2m)
Various transmission modes are available:
• Manually after pressing the PRINT key
• Continuously, according to setting
• Automatically according to stability display
• Prompted by external device
(For remote control commands see chpt. 9.8)

9.2 Pin allocation (front view)

9.3 Interface cable

38
5
1
9
6
Display unit - printer
3 (TxD)
5 (GND)
7 - 8 clench
Display unit – PC 9-channel
2 (RxD)
3 (TxD)
5 (GND)
4 - 6 clench
7 - 8 clench
Pin 2:
Pin 3:
Pin 5:
1 (RxD)
3 (GND)
3 (TxD)
2 (RxD)
5 (GND)
4 - 6 clench
7 - 8 clench
KLB-TM-BA-e-0811
Receive data
Transmit data
Signal ground