13.2 Parameter description for configuration

Parameter
Ground
setting
P 0.0
A/d
-
P 0.1
Uni
g
P 0.2 div1
0.001
P 0.3 div2
0.001
P 0.4 FulS
006.009
P 0.5 rn 2
000.000
P 0.6 Auto
0.25 d
P 0.7 wEi
000.500
P 0.8 St_u
-
P 0.9 CAL
-
P 0.A Gcor 1.00000
P 0.b
LFt
no
P 0.c
rAn
yes
P 0.d
Acu
no
P 0.E toSC
IndSt
P 0.F dFLt
-
P 0.L Line
-
KLB-TM-BA-e-0811
Range
Description
-
A/D converter value
kg - lb - g
Weighing unit
0.000 - 50
Lowest value digit and zero position of
first range
0.000 - 50
Lowest value digit and zero position of
second range
-
Maximum load
-
Switch point between first and second
range
0.10 d–
Auto zero range
5.00 d
0 - 1
Adjustment weight in respect of
maximum load
-
Initial weight for adjustment
-
Default adjustment
0.90000 –
Factor for gravity constant
1.10000
no - yes
Verified display unit
no - yes
Initial weight adjustment
yes – ON, no – OFF
no – SLA-
Selecting power supply
NiMH
No- without Bat-Lo display
SLA- battery pack 6V with Bat-Lo display
NiMH-batteries 6x 1.2V with Bat-Lo
display
IndSt-
Firmware version:
HunnA-
IndSt
Hunnb
HunnA
Hunnb
-
Reset to default setting
-
Linearization (max. 6 points)
Industrial balance
Medical balance (Hold + BMI)
Medical balance (Hold)