KERN KDE-T Operating And Installation Instructions

Display unit
Table of Contents

Advertisement

Quick Links

Operating and Installation
Instructions Display Unit
KERN KDE-T
Version 1.0
12/2012
GB
KERN & Sohn GmbH
Ziegelei 1
D-72336 Balingen
E-Mail: info@kern-sohn.com
Phone: +49-[0]7433- 9933-0
Fax: +49-[0]7433-9933-149
Internet: www.kern-sohn.com
KDE-T-BA_IA-e-1210

Advertisement

Table of Contents
loading

Summary of Contents for KERN KDE-T

  • Page 1 KERN & Sohn GmbH Ziegelei 1 Phone: +49-[0]7433- 9933-0 D-72336 Balingen Fax: +49-[0]7433-9933-149 E-Mail: info@kern-sohn.com Internet: www.kern-sohn.com Operating and Installation Instructions Display Unit KERN KDE-T Version 1.0 12/2012 KDE-T-BA_IA-e-1210...
  • Page 2: Table Of Contents

    KERN KDE-T Version 1.0 12/2012 Operating and installation instructions Display unit Contents Technical data ....................4 Appliance overview ..................5 Keyboard overview ....................6 Basic Information (General) ................7 Proper use ......................... 7 Improper Use ......................7 Warranty ........................7 Monitoring of Test Resources ..................
  • Page 3 Disposal ........................38 Instant help ....................39 Installing display unit / weighing bridge ............ 40 12.1 Technical data ......................40 12.2 Weighing system design ...................40 12.3 How to connect the platform ..................41 12.4 Configuring display devices ..................42 12.5 Service menu ......................44 KDE-T-BA_IA-e-1210...
  • Page 4: Technical Data

    1 Technical data KERN KDE-T KDE-TH Resolution 20.000 35.000 Display 6-digit Weighing Units g, kg, lb, oz Display LCD 25 mm digits with back lighting 87-1600 DMS weighing cells Sensitivity 2-3 mV/V We recommend ≥ 50 % max. Range calibration Input voltage 100 V –...
  • Page 5: Appliance Overview

    2 Appliance overview 1. Display 2. Function keys 3. RS-232 4. Input connection load cell cable 5. Mains adapter connection KDE-T-BA_IA-e-1210...
  • Page 6: Keyboard Overview

    Select menu items from menu and save your settings  Confirm the adjustment weight in the adjustment process  Menu access (longer pressing of the button)  Transfer weighing data via interface (short key press)  Taring  For plus/minus and net/total weighing: Return to weighing mode KDE-T-BA_IA-e-1210...
  • Page 7: Basic Information (General)

    The display unit may only be operated in accordance with the described default settings. Other areas of use must be released by KERN in writing. Warranty Warranty claims shall be voided in case...
  • Page 8: Monitoring Of Test Resources

    In KERN’s accredited DKD calibration laboratory test weights and display units may be calibrated (return to the national standard) fast and at moderate cost.
  • Page 9: Unpacking And Placing

    (e.g. due to mobile phones or radio equipment), static electricity accumulations or instable power supply occur. Change location or remove source of interference. Scope of delivery / serial accessories: Display Unit Mains adapter Protective cover Operating instructions KDE-T-BA_IA-e-1210...
  • Page 10: Unpacking/Installation

    Mains connection Power is supplied via the external mains adapter. The stated voltage value must be the same as the local voltage. Only use original KERN mains adapters. Using other makes requires consent by KERN. Operation using a rechargeable battery (optional) Lift-off the battery cover on the lower side of the balance.
  • Page 11: Adjustment

    The required adjustment weight depends on the capacity of the weighing system. Carry out adjustment as near as possible to the scale’s maximum weight. Info about test weights can be found on the Internet at: http://www.kern-sohn.com Observe stable environmental conditions. Stabilisation requires a certain warm-up time.
  • Page 12: Linearization

    Observe stable environmental conditions. Stabilisation requires a certain warm-up time. Do not remove adjustment weight in steps 1 to 3 during linearization, only increase it. After successful linearisation you will have to carry out calibration; see chapter “testing instruments control”. KDE-T-BA_IA-e-1210...
  • Page 13: Carry Out Linearization

    “F” will appear briefly before the weighing scale turns itself off. Linearization has now been completed successfully. We recommend that you adjust the weighing scale after linearization. (see chapter 6.6) An adjusting error or incorrect adjusting weight will be indicated by the error message; repeat adjustment procedure. KDE-T-BA_IA-e-1210...
  • Page 14: Basic Operation

    Overloading exceeding the stated maximum load (max) of the device, minus a possibly existing tare load, must be strictly avoided. This could damage the instrument. Exceeding maximum load is indicated by the display of “Error“, and an audio sound. Unload weighing system or reduce preload. KDE-T-BA_IA-e-1210...
  • Page 15: Weighing With Tare

    Put on empty vessel and press repeatedly until a flashing display of “PrEtArE” appears.  Press . The current weight on the weighing pan will be saved as pre-tare weight.  Weigh the material, the net weight will be indicated. KDE-T-BA_IA-e-1210...
  • Page 16: Plus/Minus Weighings

    „+“ and „-“. According to the same procedure also packages with the same weight can be produced, referring to a nominal weight. Back to weighing mode by pressing the button. KDE-T-BA_IA-e-1210...
  • Page 17: Parts Counting

    The balance is now in parts counting mode and counts all units on the weighing plate. The larger the reference quantity, the more accurate the parts counting. Pressing will return the weighing scale to weighing mode and show the weight for all the pieces counted. Printout example: 2pcs KDE-T-BA_IA-e-1210...
  • Page 18: Net-Total Weighings

    Weigh in component 2 ( 0.5 kg)  Press and the total weight (2.5 kg) following from component 1 and component 2 will be shown.  and the weighing scale moves to “0.0000 kg”. The memory symbol Press reappears. KDE-T-BA_IA-e-1210...
  • Page 19 Weigh in component 3 ( 5 kg)  Press and the total weight (7.5 kg) following from component 1, component 2 and component 3 will be shown. Follow the same sequence of operations for additional components. Back to weighing mode by pressing the button. KDE-T-BA_IA-e-1210...
  • Page 20: Percent Weighings

    Put test specimen on the weighing pan and the percentage in relation to the reference weight will be displayed. (example)  Press to return weighing scale to weighing mode and the weight of the test specimen will be displayed. (example) KDE-T-BA_IA-e-1210...
  • Page 21: Menu

    Confirm selected menu item with and the current setting will be shown. Change setting in selected menu item by pressing Example: ò Confirm setting Back to weighing Press to confirm required setting and the device will mode return to weighing mode. KDE-T-BA_IA-e-1210...
  • Page 22: Menu Overview

    2400 4800 Automatic shut-off function on Auto off (battery mode) Automatic shut-off function off Automatic zero tracking on Automatic zero point correction Automatic zero tracking off (zero tracking) Dependent on max load of weighing scale Selection of adjustment weight KDE-T-BA_IA-e-1210...
  • Page 23: Weighing Units (Unit)

    „Unit“ appears on the display. Press  Press , the set unit will appear on the screen. (example)  You can now use to change between individual units.  Press and the set unit will be imported. KDE-T-BA_IA-e-1210...
  • Page 24 1. Use the button to jump one digit to the left. When all the changes are ready, use the button to save this value button the „Freely selectable factor“ is taken over as and by pressing the current weighing unit. KDE-T-BA_IA-e-1210...
  • Page 25: Automatic Zero Correction (Zero Tracking)

    „Unit“ appears on the display. Press  repeatedly until “tr” appears. Press  Press , the last selected setting will be displayed. (example)  Select desired setting by  Acknowledge with . The balance changes then into weighing mode. KDE-T-BA_IA-e-1210...
  • Page 26: Selection Of Adjustment Weight

    In order to achieve high-quality weighing results in the sense of the measuring technology, it is recommended to select the nominal value as high as possible. The non delivered adjustment weights can be purchased from KERN as option.  Turn on weighing scale by and wait for zero to be displayed.
  • Page 27: Display Background Illumination

    Background illumination off Battery saving „bl“ The background illumination Battery saving will be switched off automatically 10 sec after having reached a stable weighing value.  Press to confirm the selected setting. The balance returns automatically into weighing mode. KDE-T-BA_IA-e-1210...
  • Page 28: Animal Weighing Function

    „ANL“ 10 Weighing value calculation above 10 sec. till to the value display „ANL“ 15 Weighing value calculation above 15 sec. till to the value display  Press to confirm the selected setting. The balance returns automatically into weighing mode. KDE-T-BA_IA-e-1210...
  • Page 29: Foot Switch

    8.2.6 Foot switch The foot switch is used to send the weighing values to a printer or a PC. Alternatively can also be tared. Connection and operation see in the operating instructions supplied with this foot switch. KDE-T-BA_IA-e-1210...
  • Page 30: Reset To Factory Setting

    „Unit“ appears on the display. Press  repeatedly until “rSt“ appears Press  Acknowledge using , the current setting is displayed.  Select desired setting by  Confirm you selection by pressing , the weighing scale will return to weighing mode. KDE-T-BA_IA-e-1210...
  • Page 31: Interface Parameters

    „Unit“ appears on the display. Press  repeatedly until “Pr“ appears Press  Acknowledge using , the current setting is displayed. (example)  Select desired setting by  Acknowledge selection by . The balance returns to weighing mode. KDE-T-BA_IA-e-1210...
  • Page 32: Selection Printed Edition

    (example)  Press to change setting (example)  Acknowledge with . The balance returns to weighing mode. By that way the user can configure his own data block, which then is sent to a printer or to a PC. KDE-T-BA_IA-e-1210...
  • Page 33: Baud Rate

    „Unit“ appears on the display. Press  repeatedly until “bAUd“ appears Press  Acknowledge using , the current setting is displayed. (example)  Press to change setting (example)  Acknowledge with . The balance returns to weighing mode. KDE-T-BA_IA-e-1210...
  • Page 34: Data Output Rs 232 C

    Use a suitable cable to connect the display unit to the interface of the printer. Faultless operation requires an adequate KERN interface cable. Communication parameters (Baudrate, parity) of display device and printer must match.
  • Page 35: Explanation Of The Data Transfer

    10 11 12 13 14 15 16 17 CR LF f. Format in case of fault 10 11 12 13 14 15 16 17 CR LF g. Format for unstable values for weight/quantity/percentage 10 11 12 13 14 15 16 17 CR LF KDE-T-BA_IA-e-1210...
  • Page 36: Re Cr

    13 14 15 16 17 CR LF i. Format in case of fault 10 11 12 13 14 15 16 17 CR LF j. Format for unstable values for weight/quantity/percentage 10 11 12 13 14 15 16 17 CR LF KDE-T-BA_IA-e-1210...
  • Page 37: Output On Bar Code Printer

    The printer format is stored in the printer, i.e. in case of a failure the printer cannot be changed with a new one from factory, previously it is necessary that KERN installs the respective software. The Zebra printer and the balance must be connected to the delivered interface cable when they are switched off.
  • Page 38: Servicing, Maintenance, Disposal

    Spilled weighing goods must be removed immediately. 10.2 Servicing, maintenance The appliance may only be opened by trained service technicians who are authorized by KERN. Before opening, disconnect from power supply. 10.3 Disposal Disposal of packaging and appliance must be carried out by operator according to valid national or regional law of the location where the appliance is used.
  • Page 39: Instant Help

    Great fluctuations in temperature. Warm-up time was ignored. Electromagnetic fields / static charging (choose different location/switch off interfering device if possible) Should other error messages occur, switch device off and then on again. If the error message remains inform manufacturer. KDE-T-BA_IA-e-1210...
  • Page 40: Installing Display Unit / Weighing Bridge

    4 % of the scale’s capacity. The addition of weighing scales capacity, preload and the total zero setting range give the required capacity for the weighing cell. To avoid overloading of the weighing cell, include an additional safety margin. Smallest desired display division KDE-T-BA_IA-e-1210...
  • Page 41: How To Connect The Platform

    Disconnect the display unit from the power supply.  Weld the individual wires of the load cell cable to the printed circuit board.  Pin assignment is shown on the diagram below. Load cell Display green Sig+ green Sen+ white white Sig- black black Sens- KDE-T-BA_IA-e-1210...
  • Page 42: Configuring Display Devices

    Release both keys. “SCL.tYP” will be displayed followed by “1rAnGE“ Select menu items With help of , the individual menu items can be selected one after the other. ò Change settings Press to change settings Confirm setting to acknowledge settings KDE-T-BA_IA-e-1210...
  • Page 43 Readability for first weighing range selectable, dependent on max load SCL.CAP 2 Balance capacity (Max) 2nd weighing range rESOLE 2 Readability for 2nd weighing range selectable, dependent on max load After configuration linearize the weighing system Linearization (See chap. 6.7.1) KDE-T-BA_IA-e-1210...
  • Page 44: Service Menu

    A numeric value appears and the digit on the right is flashing.  Press to increase the numeric value or (example) decrease. Press to change to a different decimal place. Set weighing scale capacity (here is an example of (example) 50 kg) and press to confirm. KDE-T-BA_IA-e-1210...
  • Page 45: Kde-T-Ba_Ia-E

    “P1.0x” is displayed. Release both keys. “SCL.TYP”will appear briefly followed by “1.rAnGE“.  Setting single range (“1.rAnGE“) – or dual range (“2.rAnGE“) weighing scale   to change between “1.rAnGE”and “2.rAnGE”. Press  Select „2.rAnGE“ and confirm by KDE-T-BA_IA-e-1210...
  • Page 46 “rESOLE 2”will appear briefly followed by a numeric value. As above, enter required readability and press to confirm the required value. “F” will appear briefly before the weighing scale changes to adjusting mode and “CAL” will be displayed.   Linearizing a weighing scale (see chap 6.7.1) KDE-T-BA_IA-e-1210...