7 Data output RS 232C

You can print weighing data automatically via the RS 232C interface or manually by pressing
via the interface according to the setting in the menu.
This data exchange is asynchronous using ASCII - Code.
The following conditions must be met to provide successful communication between the
weighing system and the printer.
 Use a suitable cable to connect the display unit to the interface of the printer.
Faultless operation requires an adequate KERN interface cable.
 Communication parameters (baud rate, bits and parity) of display unit and printer
must match. For a detailed description of interface parameters see chap. Fehler!
Verweisquelle konnte nicht gefunden werden., menu block „P1 COM" or ‚"P2
COM"

7.1 Technical data

Connection
9 pin d-subminiature bushing
Baud rate
Optional 600/1200/2400/4800/9600
8 bits, no parity / 7 bits, even parity / 7 bits, odd parity
Parity

7.2 Printer mode

Printout examples (KERN YKB-01N):
Weighing
ST, GS
- 19 -
1.000kg
Pin 2 input
Pin 3 output
Pin 5 signal earth
KFB-TM/KFN-TM-SH-e-1410