Factor For Constant Of Gravitation P 0.A Gcor - KERN KLB-TM Operating Manual

Table of Contents

Advertisement

For approx. 2 seconds appears „LOAD" on the display, after that the defined value
of the correction point. With unloaded weighing plate, 0.000kg is shown in the
display.
First look for that no load is on the weighing plate and the following symbols can be
seen ( a,
), if not, press the a-button.
Now put a defined weight on the weighing plate.
After the weight display having been stabilized, press the PRINT-button in order to
reach the next step.
After the stabilization of the weight display at some models the next step is
automatically called up.
The message „nEHt ?" appears on the display. Now you must make up your
decision if you want further linearization points.
If you wish further linearization points, unload the weighing plate and subsequently
press the PRINT-button, „Pnt2" appears on the display (second linearization point).
Now you must procede here as described aboce in „Pnt1".
The program allows to define up to 6 linearisation points. After having set the last
point (the 6th one), „DonE" appears on the display and subsequently returns to the
menu, on the display appears 0.L LinE .
If less than 6 linearization points are necessary, press the F-button when the
message „nEHt ?" appears, in order to finish the function and to return to the
selection menu.
In the display appears „donE" and then 0.L LinE .
Jumping back to the weighing mode saves the settings made.
Notice:
If after pressing the PRINT-button appears the „-LO-" display for approx 2 seconds,
followed by „donE" and after that 0.L LinE, the function of the linearization
correction is locked.

13.5 Factor for constant of gravitation P 0.A Gcor

This correction factor is used to adapt the constant of gravitation for locations at
different degrees of latitude.
This makes it possible to carry out adjustment at a different location than that where
the display unit with connected weighing cell will later be operated.
This factor is computed by applying the formula below:
g (usage) / g (adjustment location) = g_cor
Allowable values range from 0.90000 to 1.10000.
Notice: Is the display unit with connected weighing cell operated at the place of
adjustment, this factor is to be set at 1.00000.
52
KLB-TM-BA-e-0811