Operating Manual
Display units
KERN KLB-TM
Version 1.1
01/2008
GB
KERN & Sohn GmbH
Ziegelei 1
D-72336 Balingen
E-Mail: info@kern-sohn.com
Tel: +49-[0]7433- 9933-0
Fax: +49-[0]7433-9933-149
Internet: www.kern-sohn.com
KLB-TM-BA-e-0811

Advertisement

Table of Contents
loading

Summary of Contents for KERN KLB-TM

  • Page 1 KERN & Sohn GmbH Ziegelei 1 Tel: +49-[0]7433- 9933-0 D-72336 Balingen Fax: +49-[0]7433-9933-149 E-Mail: info@kern-sohn.com Internet: www.kern-sohn.com Operating Manual Display units KERN KLB-TM Version 1.1 01/2008 KLB-TM-BA-e-0811...
  • Page 2: Table Of Contents

    KERN KLB-TM Version 1.1 01/2008 Operating Manual Display Unit Table of Contents Technical data __________________________________________________ 5 Declaration of conformity ________________________________________ 6 Basic Information (General)_______________________________________ 8 Proper use _______________________________________________________ 8 Improper Use ____________________________________________________ 8 Warranty ________________________________________________________ 8 Monitoring of Test Resources _______________________________________ 9...
  • Page 3 Baud rate for RS232 interface ______________________________________ 42 Communication protocol / remote control commands __________________ 43 9.8.1 Response messages from balance _______________________________________ 43 9.8.2 Manual output________________________________________________________ 44 9.8.3 Continuous output ____________________________________________________ 45 Error messages ______________________________________________ 46 Service, maintenance, disposal _________________________________ 46 11.1 Cleaning _______________________________________________________ 46 KLB-TM-BA-e-0811...
  • Page 4 Default adjustment parameter P 0.9 CAL _____________________________ 51 13.4.1 Linearization Parameter P 0.L LinE _______________________________________ 51 13.5 Factor for constant of gravitation P 0.A Gcor _________________________ 52 13.6 Calibrated display unit P 0.b LFt ____________________________________ 53 13.7 Selecting the power supply P 0.d Acu _______________________________ 53 KLB-TM-BA-e-0811...
  • Page 5: Technical Data

    1 Technical data KERN KLB-TM Display 6-digit Resolution 6000 e / 100000 d verifiable / non-verifiable Weighing ranges Display LCD numeral height 18 mm, backlit Nominal load 1 …50.000 Warm-up time 2 hours Battery power supply Standard Weighing cell connection...
  • Page 6: Declaration Of Conformity

    EN 60950 :2000/A11 :2000 Low Voltage 89/336/EEC EN61000-4-2 :1999 EN 61000-4-3 :1996 EN 61000-4-4 :1999 EN 61000-4-11 :1997 Date: 01.02.2007 Signature: Gottl. KERN & Sohn GmbH Management Gottl. KERN & Sohn GmbH, Ziegelei 1, D-72336 Balingen, Tel. +49-[0]7433/9933-0,Fax +49-[0]7433/9933-149 KLB-TM-BA-e-0811...
  • Page 7 KLB-TM terminals in connection with approved load cells nur gültig für KLB-TM Terminals in Verbindung mit zugelassenen Lastzellen valable uniquement pour les terminaux KLB-TM en liaison avec des cellules de charge homologuées sólo válido para terminales KLB-TM en combinación con células de carga aprobadas valido solo per terminali KLB-TM in collegamento con celle di carico approvate Date: 01.02.2007...
  • Page 8: Basic Information (General)

    The display unit may only be operated in accordance with the described default settings. Other areas of use must be released by KERN in writing. 3.3 Warranty Warranty claims shall be voided in case •...
  • Page 9: Monitoring Of Test Resources

    KERN’s home page (www.kern-sohn.com with regard to the monitoring of display units’ test substances and the test weights required for this. In KERN’s accredited DKD calibration laboratory test weights and display units may be calibrated (return to the national standard) fast and at moderate cost.
  • Page 10: Unpacking, Setup And Commissioning

    Carefully take the display unit out of the packaging, remove the plastic jacket and mount the display unit at the intended workplace. 6.2.1 Setup Mount the display unit in a way that facilitates operation and where it is easy to see. KLB-TM-BA-e-0811...
  • Page 11: Scope Of Delivery

    Before connecting or disconnecting of additional devices (printer, PC) to the data interface, always disconnect the display unit from the power supply. With your display unit, only use accessories and peripheral devices by KERN, as they are ideally tuned to your display unit.
  • Page 12: Adjustment

    Place the required calibration weight carefully in the centre of the weighing plate and press the PRINT key. CAL appears on the display, calibration is started. unLoAd unLoAd appears on the display, calibration is complete DonE Remove the test weights 6.2.uCAL KLB-TM-BA-e-0811...
  • Page 13 The display unit returns to weighing mode. Message Err 4 appears for a calibration error or incorrect test weight. In this case, repeat the calibration sequence. Keep the test weight with the weighing plate. Daily control of the weighing exactness is recommended for quality-relevant applications. KLB-TM-BA-e-0811...
  • Page 14: Operation

    The display unit carries out an automatic check, when started. First this involves all display elements being lit up. Next the software version appears. After the message –test- , nlnnH will appear which indicates the use of chargeable batteries. KLB-TM-BA-e-0811...
  • Page 15: Overview Of Display

    0.00 • • • +/- tolerance with respect to reference weight within prescribed limits 0.00 • • • +/- tolerance with respect to reference weight entry of upper limit or weight above tolerance 0.00 • KLB-TM-BA-e-0811...
  • Page 16: User Menu

    Storing maximum value of weight 4.9 Add Add function P5 othr 5.1 bL Auto Background illumination 5.2 blbA Intensity of backlighting 5.3 bEEP Key sound 5.4 t1 Automatic disconnection 5.5 CHr6 Disconnection battery control P6 CAL 6.1 St_u not documented 6.2 uCal Adjustment KLB-TM-BA-e-0811...
  • Page 17: Keyboard Overview In Menu

    For this, press the F key repeatedly until SauE ? appears on the display. Any changes carried out are stored by pressing the PRINT key. To cancel changes, press the F key. Afterwards the balance automatically jumps back to weighing mode. KLB-TM-BA-e-0811...
  • Page 18: Weighing

    To change your selection, press the a key. P3.1.StUn Setting will be imported by pressing the PRINT key. For this, press the F key repeatedly until SavE ? appears on the display. Store changes carried out by pressing the PRINT key. KLB-TM-BA-e-0811...
  • Page 19: Weighing Units Switch-Over

    Setting will be imported by pressing the PRINT key. Selection possibility: The options available depend on the setting made during the configuration of the connected weighing plate. You may switch either between kg, lb and N or between g and ct. KLB-TM-BA-e-0811...
  • Page 20: Taring

    (active digit flashing) Setting is imported by pressing the PRINT key and 0.00 g balance returns to weighing mode. This type of taring may be carried out in weighing mode at any time. KLB-TM-BA-e-0811...
  • Page 21: P4 Func - Operating Modes

    Percentage display via Prcb Press the a key manual default AtAr Press the a key Auto-taring Press the a key Peak value function Press the a key Adding Confirm settings by pressing the PRINT key. Balance jumps back to sub-menu 4.1.FFun. KLB-TM-BA-e-0811...
  • Page 22: Settings For Menu Item P4.1 Ffun For "All

    Please repeat this sequence of operations for any other operating mode available. 4.2.Funi 4.3.PcS 4.4.HiLo 4.5 PrcA 4.6 Prcb 4.7 AtAr 4.8 toP 4.9 Add For this, press the F key repeatedly until SavE ? appears on the display. Store changes carried out by pressing the PRINT key. KLB-TM-BA-e-0811...
  • Page 23: P5 Othr - Operating Parameter

    To change your selection, press the a key. 5.1.bl Setting will be imported by pressing the PRINT key. For this, press the F key repeatedly until SavE ? appears on the display. Store changes carried out by pressing the PRINT key. KLB-TM-BA-e-0811...
  • Page 24: Intensity Of Background Lighting For Battery Operation

    To change your selection, press the a key. 5.2.blbA Setting will be imported by pressing the PRINT key. For this, press the F key repeatedly until SavE ? appears on the display. Store changes carried out by pressing the PRINT key. KLB-TM-BA-e-0811...
  • Page 25: Keyboard Sound For Pressed Key

    Change the battery immediately if the battery symbol lights up during battery operation or charge the accumulator during accumulator operation or the balance will be turned off shortly. This symbol will be flashing whilst the battery is being loaded. KLB-TM-BA-e-0811...
  • Page 26: Battery Status Display

    Press TARE key Repeatedly press the a key until the following 5.5. CHr6 display appears To change your selection, press the TARE key. To change your selection, press the a key. 5.5.CHr6 Setting will be imported by pressing the PRINT key. KLB-TM-BA-e-0811...
  • Page 27: P1 Read - Basic Settings

    Robust but slow (very busy set-up location) Operation: Display: P1 rEAd Selecting the menu item 1.1.Fil Press TARE key Press TARE key To change your selection, press the a key. 1.1.Fil Setting will be imported by pressing the PRINT key. KLB-TM-BA-e-0811...
  • Page 28: Auto Zero

    Press TARE key Press the a key until the following display 1.2 Auto appears To change your selection, press the TARE key. To change your selection, press the a key. 1.2 Auto Setting will be imported by pressing the PRINT key. KLB-TM-BA-e-0811...
  • Page 29: Tare Function

    1.3.tArA display appears AtAr To change your selection, press the TARE key. To change your selection, press the a key. tArF To change your selection, press the a key. 1.3.tArA Setting will be imported by pressing the PRINT key. KLB-TM-BA-e-0811...
  • Page 30: Median Filter

    To change your selection, press the a key. 1.4.tArA Setting will be imported by pressing the PRINT key. For this, press the F key repeatedly until SavE ? appears on the display. Store changes carried out by pressing the PRINT key. KLB-TM-BA-e-0811...
  • Page 31: Parts Counting

    Operate the PRINT key 00000 Use the TARE key, to select the digit to be changed, the a key is used to increase the numerical value (active digit flashing) 00015 Confirm entered reference number of items by pressing the PRINT key. LoAD KLB-TM-BA-e-0811...
  • Page 32: Percent Determination

    Press PRINT key, the weight is taken as reference 100.00 % (100%). Now you can place the test objects onto the weighing 70.37 % plate; the percentage is displayed Go back to weighing mode by pressing the F key twice. 0.00 g KLB-TM-BA-e-0811...
  • Page 33: Calculate The Reference Weight By Numeric Input

    Press PRINT key, the weight is taken as reference 0.00 % (100%). Now you can place the test objects onto the weighing 70.37 % plate; the percentage is displayed Go back to weighing mode by pressing the F key twice. 0.00 g KLB-TM-BA-e-0811...
  • Page 34: Tolerance Weighing

    Go back to weighing mode by pressing the F key twice. 0.00 g The value entered for the upper tolerance limit must be greater than the lower tolerance value. Error message –Lo- appears on the display and the display of the unit returns to weighing mode. KLB-TM-BA-e-0811...
  • Page 35: Automatic Taring

    Go back to weighing mode by pressing the F key twice. 0.00 g Important: Ensure that the minimum weight entered (For setting see chpt. 9.5) is lower than the weight of the weighing container or the weighing container will not be tared automatically. KLB-TM-BA-e-0811...
  • Page 36: Storing Maximum Value Of Weight

    Place weight on weighing plate. 234.56 g Set up additional weight 234.56 g Remove 1 weight 234.56 g Remove all weights 0.00 g To reset, press the a key Go back to weighing mode by pressing the F key twice. 0.00 g KLB-TM-BA-e-0811...
  • Page 37: Summation Of Weighing Values

    To achieve this, call the function as described above and the sum so far will appear automatically. If the display range of the display unit is exceeded by the sum total, “5-Full” will appear. KLB-TM-BA-e-0811...
  • Page 38: Data Output Rs 232 C

    Baud rate selectable at 2400, 4800, 9600, 19200 and 38400 Baud Miniature plug-in necessary (9 pole D-Sub) For operation with interface faultless operation is only ensured with the correct KERN – interface cable (max. 2m) Various transmission modes are available: • Manually after pressing the PRINT key •...
  • Page 39: Description Of Data Output Type

    (PRINT key) StAb Data output for stable weighing value (PRINT key) rEPL Automatic output function (See chpt. 9.5) CntA Continuous output in standard weighing unit Cntb Continuous output in current weighing unit KLB-TM-BA-e-0811...
  • Page 40: Input Minimum Weight For Some Functions

    (active digit flashing) Press the PRINT key; the minimum weight will be 2.2. S_Lo imported. For this, press the F key repeatedly until SavE ? appears on the display. Store changes carried out by pressing the PRINT key. KLB-TM-BA-e-0811...
  • Page 41: Baud Rate For Rs232 Interface

    To change your selection, press the a key. 9600 2.3. bAud Press the PRINT key; the baud rate will be imported. For this, press the F key repeatedly until SavE ? appears on the display. Store changes carried out by pressing the PRINT key. KLB-TM-BA-e-0811...
  • Page 42: Baud Rate For Rs232 Interface

    8d1SnP : 8 data bit, 1 stop bit, no parity 8d2SnP : 8 data bit, 2 stop bit, no parity 8d1SEP : 8 data bit, 1 stop bit, EVEN parity 8d1SoP : 8 data bit, 1 stop bit, ODD parity KLB-TM-BA-e-0811...
  • Page 43: Communication Protocol / Remote Control Commands

    XX_ I CR LF Instruction received; impossible to carry out XX_^ CR LF Instruction received but time overflow error occurred XX_v CR LF Instruction received, but insufficient load XX_E CR LF Error during execution, timeout for stable weighing value exceeded KLB-TM-BA-e-0811...
  • Page 44: Manual Output

    Weight Space Unit indicator nal sign Stability indicator Space if stable, ? if not stable ^ if overload v if underload Presign: Space, if positive negative sign, if negative Weight: 9 signs, right justified Unit: 3 signs, left justified KLB-TM-BA-e-0811...
  • Page 45: Continuous Output

    Weig Spac Unit indicator tional sign Stability indicator Space if stable, ? if not stable ^ if overload v if underload Presign: Space, if positive negative sign, if negative Weight: 9 signs, right justified Unit: 3 signs, left justified KLB-TM-BA-e-0811...
  • Page 46: Error Messages

    Spilled weighing goods must be removed immediately. 11.2 Service, maintenance The appliance may only be opened by trained service technicians who are authorized by KERN. Before opening, disconnect from power supply. 11.3 Disposal Disposal of packaging and appliance must be carried out by operator according to valid national or regional law of the location where the appliance is used.
  • Page 47: Instant Help

    • Warm-up time was ignored. • Electromagnetic fields / static charging (choose different location/switch off interfering device if possible) Should other error messages occur, switch display unit off and then on again. If the error message remains inform manufacturer. KLB-TM-BA-e-0811...
  • Page 48: Connecting A Weighing Cell To The Display Unit

    + REF For 4-wire cell, close JP1, otherwise – reference signal Connect cable shield of weighing cell unless shield is already connected to weighing cell. If the shield is already connected to the weighing cell, this terminal remains empty. KLB-TM-BA-e-0811...
  • Page 49: Parameter Description For Configuration

    NiMH-batteries 6x 1.2V with Bat-Lo display P 0.E toSC IndSt IndSt- Firmware version: HunnA- IndSt Industrial balance Hunnb HunnA Medical balance (Hold + BMI) Hunnb Medical balance (Hold) P 0.F dFLt Reset to default setting P 0.L Line Linearization (max. 6 points) KLB-TM-BA-e-0811...
  • Page 50: Calling Menu For Entering Parameter Of Configuration

    To store, press the PRINT key; to exit without storing, press the F key. Notice: After making and storing changes, turn the display unit off, then turn it on again. Micro switch S1 After carrying out the configuration, close the case and insert the screws. KLB-TM-BA-e-0811...
  • Page 51: Default Adjustment Parameter P 0.9 Cal

    When pressing the PRINT button, „0“00.000 kg appears on the display. The first number flashes and with the help of the TARE and a-buttons the respective position and the value can be selected, the first linearization point is taken over using the PRINT button. KLB-TM-BA-e-0811...
  • Page 52: Factor For Constant Of Gravitation P 0.A Gcor

    This factor is computed by applying the formula below: g (usage) / g (adjustment location) = g_cor Allowable values range from 0.90000 to 1.10000. Notice: Is the display unit with connected weighing cell operated at the place of adjustment, this factor is to be set at 1.00000. KLB-TM-BA-e-0811...
  • Page 53: Calibrated Display Unit P 0.B Lft

    All voltage limits for the various storage batteries are determined by setting this parameter. This parameter is ineffective if the display unit is powered by the plug-in power unit. KLB-TM-BA-e-0811...