Setting Of Data Transfer Mode; Setting Of Baud Rate; Data Output Rs 232 C; Technical Data - KERN YKD Series Operating Instructions Manual

Second display
Table of Contents

Advertisement

6.5 Setting of data transfer mode

Setting of data transfer mode
Press the Mode 1 key and keep it pressed down. Additionally
press the Print1 key and keep it pressed down until "Pr 1"
appears. Release both keys.
Press the Mode1 key, e.g. "Pr PC" appears.
The mode can be converted using the Print1 button
(Pr PC; AU PC; AU Pr ; re Cr ; bA Pr ).
The changed setting is taken over by pressing the Mode1
button.
Display returns to display mode.

6.6 Setting of baud rate

Setting of baud rate
Press the Mode 1 key and keep it pressed down. Additionally
press the Print1 key and keep it pressed down until "Pr 1"
appears. Release both keys.
Press Print1 button, „baud" appears.
Use the Mode1 key, to confirm.
The baud rate can be changed using the Print1 button (1200,
2400, 4800, 9600, 19200, 38400).
The changed setting is taken over by pressing the Mode1
button.
Display returns to display mode.

7 Data output RS 232 C

7.1 Technical data

8-bit ASCII Code
1 start bit, 8 data bits, 1 stop bit, no parity bit
Baud rate selectable at 1200, 2400, 4800, 9600 and 19200 Baud
Miniature plug-in necessary (9 pole D-Sub)
For operation with interface faultless operation is only ensured with the correct
KERN – interface cable (max. 2m)
10
Balance display
Pr 1
Pr PC
AU Pr
0.0 g
Balance display
Pr 1
bAUd 1
4800
9600
0.0 g
YKD-BA-e-0710