KERN KMA-TM Operating And Installation Instructions

KERN KMA-TM Operating And Installation Instructions

Display unit

Advertisement

Quick Links

Operating and Installation
Instructions
Display Unit
KERN KMA-TM
Version 1.0
05/2013
GB
KERN & Sohn GmbH
Ziegelei 1
D-72336 Balingen
E-Mail: info@kern-sohn.com
Phone: +49-[0]7433- 9933-0
Fax: +49-[0]7433-9933-149
Internet: www.kern-sohn.com
KMA-TM-BA_IA-e-1310

Advertisement

Table of Contents
loading

Summary of Contents for KERN KMA-TM

  • Page 1 KERN & Sohn GmbH Ziegelei 1 Phone: +49-[0]7433- 9933-0 D-72336 Balingen Fax: +49-[0]7433-9933-149 E-Mail: info@kern-sohn.com Internet: www.kern-sohn.com Operating and Installation Instructions Display Unit KERN KMA-TM Version 1.0 05/2013 KMA-TM-BA_IA-e-1310...
  • Page 2: Table Of Contents

    KERN KMA-TM Version 1.0 05/2013 Operating and installation instructions Display unit Contents Technical data ....................4 Appliance overview ..................5 Overview of display ....................6 Keyboard overview ....................7 Basic instructions ..................8 Proper use ......................... 8 Improper Use ......................8 Warranty ........................
  • Page 3 Configure display unit ....................22 11.4.1 Navigation in the menu ......................22 11.4.2 Menu overview ........................23 11.4.3 Carry out configuration ......................24 Verification ....................27 Adjustment ....................29 Declaration of Conformity / Test Certificate / Type Approval Certificate Fehler! Textmarke nicht definiert. KMA-TM-BA_IA-e-1310...
  • Page 4: Technical Data

    195 x 118 x 83 (B x D x H) mm Admissible -10°C – 40°C ambient temperature Net weight 1 kg Medical product in Class I when using as weighing system KERN MCC or accordance with KERN MPC 93/42/EEC KMA-TM-BA_IA-e-1310...
  • Page 5: Appliance Overview

    2 Appliance overview 1. Weight display 2. Battery charge status 3. Keyboard 4. Weighing unit 5. Adjustment switch 6. Terminal power supply unit 7. Connection connecting cable „display unit - loading cell" 8. Battery compartment KMA-TM-BA_IA-e-1310...
  • Page 6: Overview Of Display

    HOLD function active BMI function Illuminated while BMI function is enabled Lights when the voltage drops below the prescribed minimum. Rechargeable battery symbol Lights when the rechargeable battery capacity is almost exhausted. Lights when the rechargeable battery is fully charged. KMA-TM-BA_IA-e-1310...
  • Page 7: Keyboard Overview

    In menu: • Call up menu • Select menu items For numeric entry: • Increase numerical value Calculation of the Body Mass Index In menu: • Confirm selection For numeric entry: • Confirm numerical value Data-Hold function Turn on/off KMA-TM-BA_IA-e-1310...
  • Page 8: Basic Instructions

    The display unit may only be operated in accordance with the described default settings. Other areas of use must be released by KERN in writing. 3.3 Warranty Warranty claims shall be voided in case •...
  • Page 9: Monitoring Of Test Resources

    KERN’s home page (www.kern-sohn.com with regard to the monitoring of display units’ test substances and the test weights required for this. In KERN’s accredited DKD calibration laboratory test weights and display units may be calibrated (return to the national standard) fast and at moderate cost.
  • Page 10: Unpacking, Setup And Commissioning

    Carefully remove the display unit from packaging, remove plastic cover and place it in the designated work area. and place it in the designated work area. 6.3 Scope of delivery  Display Unit  Mains adapter  Operating instructions KMA-TM-BA_IA-e-1310...
  • Page 11: Mains Connection

    Power is supplied via the external mains adapter. The stated voltage value must be the same as the local voltage. Only use original KERN mains adapters. Using other makes requires consent by KERN. The small sticker attached to the side of the display unit indicates the power port: The LED remains illuminated as long as the weighing scale remains connected to the mains.
  • Page 12: Rechargeable Battery Operation

    (mains, accumulator or battery) and be switched on. The accuracy of the balance depends on the local acceleration of gravity. The value of gravity acceleration is shown on the type plate. KMA-TM-BA_IA-e-1310...
  • Page 13: Operation

     The taring process can be repeated any number of times. The limit is reached when the taring range capacity (see type plate) is full.  To delete the tare value, remove load from balance and press KMA-TM-BA_IA-e-1310...
  • Page 14: Hold Function

    The „HOLD“ symbol disappears. 7.7 See second decimal point (not verified value) Press and hold for about 2 s whilst weighed result is being shown. The second decimal place will be shown for approx. 5 s. KMA-TM-BA_IA-e-1310...
  • Page 15: Calculation Of The Body Mass Index

    < 18.5 Normal weight 18.5 – 24.9 Average Overweight > 25.0 Pre-adipose 25.0 – 29.9 A bit high Adipose degree I 30.0 – 34.9 Increased Adipose degree II 35.0 – 39.9 High Adipose degree III > 40 Very high KMA-TM-BA_IA-e-1310...
  • Page 16: Automatic Switch-Off Function „Auto Off

    AUTO-OFF after 5 minutes without changing the weight [oFF 15] AUTO-OFF after 15 minutes without changing the weight [oFF 30] AUTO-OFF after 30 minutes without changing the weight  Confirm with , [F1 oFF] will be displayed  Return to weighing mode using KMA-TM-BA_IA-e-1310...
  • Page 17: Display Background Illumination

    Continuous background lighting bL off Background illumination off bL AU Automatic background illumination on when weighing surface is loaded or key pressed.  Confirm with , [F3 bk] will be displayed.  Return to weighing mode using KMA-TM-BA_IA-e-1310...
  • Page 18: Error Messages

     Do not use aggressive detergents (solvents or similar). 9.2 Servicing, maintenance The appliance may only be opened by trained service technicians who are authorized by KERN. Disconnect the device from mains before opening. 9.3 Disposal Disposal of packaging and appliance must be carried out by operator according to valid national or regional law of the location where the appliance is used.
  • Page 19: Instant Help

     Weighing system is not levelled.  Electromagnetic fields / static charging (choose different location/switch off interfering device if possible) Should other error messages occur, switch balance off and then on again. If the error message remains inform manufacturer. KMA-TM-BA_IA-e-1310...
  • Page 20: Installation Display Unit

    The addition of weighing scales capacity, preload and the total zero setting range give the required capacity for the weighing cell. To avoid overloading of the weighing cell, include an additional safety margin. • Smallest desired display division • Verifiable, if required KMA-TM-BA_IA-e-1310...
  • Page 21: How To Connect The Platform

    11.3 How to connect the platform  Disconnect the display unit from the power supply.  Solder the individual leads of the load cell cable onto the circuit board, see diagrams below. Weighing cell (Zemic L6D) Display Unit KERN KMA-TM KMA-TM-BA_IA-e-1310...
  • Page 22: Configure Display Unit

     Confirm selected function by . The current setting will be displayed.  Select desired setting by and confirm with , the balance returns to the menu. Exit menu/  Press , the balance will return to weighing mode. Return to weighing mode KMA-TM-BA_IA-e-1310...
  • Page 23: Menu Overview

    1, div 2, div 5, div 10, Readability, s div 20, div 50 Balance capacity (/max) Adjustment, see chap. 16.1 P3 Pro tri* Not documented CoUnt Internal A/D converter value rESEt Reset weighing scale to factory setting SEtGrA Not documented * default setting KMA-TM-BA_IA-e-1310...
  • Page 24: Carry Out Configuration

    , the currently set position of the decimal dot is displayed. Press to select the desired setting. Selectable 0, 0.0, 0.00, 0.000, 0.0000. Confirm input with , the balance returns to the menu.  Use to select the next menu item [inC]. KMA-TM-BA_IA-e-1310...
  • Page 25  Wait for stability display "STABLE", then confirm with  The size of the currently set adjustment weight will be displayed. To change, select the digit to be altered by , and the (example) numerical value by  Confirm with , [LoAd] appears. KMA-TM-BA_IA-e-1310...
  • Page 26 An adjusting error or incorrect adjusting weight will be indicated by the error message; repeat adjustment procedure. An adjusting error or incorrect adjustment weight will generate an error message („Err 4“), repeat the adjustment process. KMA-TM-BA_IA-e-1310...
  • Page 27: Verification

    The weighing result of the balance is outside the error limit. Therefore, in • regular intervals load balance with known test weight (ca. 1/3 of the max. load) and compare with displayed value. The reverification deadline has been exceeded. • KMA-TM-BA_IA-e-1310...
  • Page 28 Position adjustment switch and seals: Self-destroying seal mark Cover Adjustment switch KMA-TM-BA_IA-e-1310...
  • Page 29: Adjustment

    Prepare the required adjustment weight. The weight to be used depends on the capacity of the scale. Carry out adjustment as near as possible to the scale’s maximum weight. Info about test weights can be found on the Internet at: http://www.kern-sohn.com. • Observe stable environmental conditions. Stabilisation requires a certain warm-up time.
  • Page 30  The size of the currently set adjustment weight will be displayed. The active digit is flashing. To change, select the digit to be altered by , and the (example) numerical value by  Confirm with , [LoAd] appears. KMA-TM-BA_IA-e-1310...
  • Page 31 An adjusting error or incorrect adjustment weight will generate an error message („Err 4“), repeat the adjustment process. 14 Enclosure Type Approval Certificate when using as weighing system KERN MCC or MPC KMA-TM-BA_IA-e-1310...