KERN KMA-TM Version 1.0 05/2013 Operating and installation instructions Display unit Contents Technical data ....................4 Appliance overview ..................5 Overview of display ....................6 Keyboard overview ....................7 Basic instructions ..................8 Proper use ......................... 8 Improper Use ......................8 Warranty ........................
Page 3
Configure display unit ....................22 11.4.1 Navigation in the menu ......................22 11.4.2 Menu overview ........................23 11.4.3 Carry out configuration ......................24 Verification ....................27 Adjustment ....................29 Declaration of Conformity / Test Certificate / Type Approval Certificate Fehler! Textmarke nicht definiert. KMA-TM-BA_IA-e-1310...
195 x 118 x 83 (B x D x H) mm Admissible -10°C – 40°C ambient temperature Net weight 1 kg Medical product in Class I when using as weighing system KERN MCC or accordance with KERN MPC 93/42/EEC KMA-TM-BA_IA-e-1310...
HOLD function active BMI function Illuminated while BMI function is enabled Lights when the voltage drops below the prescribed minimum. Rechargeable battery symbol Lights when the rechargeable battery capacity is almost exhausted. Lights when the rechargeable battery is fully charged. KMA-TM-BA_IA-e-1310...
In menu: • Call up menu • Select menu items For numeric entry: • Increase numerical value Calculation of the Body Mass Index In menu: • Confirm selection For numeric entry: • Confirm numerical value Data-Hold function Turn on/off KMA-TM-BA_IA-e-1310...
The display unit may only be operated in accordance with the described default settings. Other areas of use must be released by KERN in writing. 3.3 Warranty Warranty claims shall be voided in case •...
KERN’s home page (www.kern-sohn.com with regard to the monitoring of display units’ test substances and the test weights required for this. In KERN’s accredited DKD calibration laboratory test weights and display units may be calibrated (return to the national standard) fast and at moderate cost.
Carefully remove the display unit from packaging, remove plastic cover and place it in the designated work area. and place it in the designated work area. 6.3 Scope of delivery Display Unit Mains adapter Operating instructions KMA-TM-BA_IA-e-1310...
Power is supplied via the external mains adapter. The stated voltage value must be the same as the local voltage. Only use original KERN mains adapters. Using other makes requires consent by KERN. The small sticker attached to the side of the display unit indicates the power port: The LED remains illuminated as long as the weighing scale remains connected to the mains.
(mains, accumulator or battery) and be switched on. The accuracy of the balance depends on the local acceleration of gravity. The value of gravity acceleration is shown on the type plate. KMA-TM-BA_IA-e-1310...
The taring process can be repeated any number of times. The limit is reached when the taring range capacity (see type plate) is full. To delete the tare value, remove load from balance and press KMA-TM-BA_IA-e-1310...
The „HOLD“ symbol disappears. 7.7 See second decimal point (not verified value) Press and hold for about 2 s whilst weighed result is being shown. The second decimal place will be shown for approx. 5 s. KMA-TM-BA_IA-e-1310...
< 18.5 Normal weight 18.5 – 24.9 Average Overweight > 25.0 Pre-adipose 25.0 – 29.9 A bit high Adipose degree I 30.0 – 34.9 Increased Adipose degree II 35.0 – 39.9 High Adipose degree III > 40 Very high KMA-TM-BA_IA-e-1310...
AUTO-OFF after 5 minutes without changing the weight [oFF 15] AUTO-OFF after 15 minutes without changing the weight [oFF 30] AUTO-OFF after 30 minutes without changing the weight Confirm with , [F1 oFF] will be displayed Return to weighing mode using KMA-TM-BA_IA-e-1310...
Continuous background lighting bL off Background illumination off bL AU Automatic background illumination on when weighing surface is loaded or key pressed. Confirm with , [F3 bk] will be displayed. Return to weighing mode using KMA-TM-BA_IA-e-1310...
Do not use aggressive detergents (solvents or similar). 9.2 Servicing, maintenance The appliance may only be opened by trained service technicians who are authorized by KERN. Disconnect the device from mains before opening. 9.3 Disposal Disposal of packaging and appliance must be carried out by operator according to valid national or regional law of the location where the appliance is used.
Weighing system is not levelled. Electromagnetic fields / static charging (choose different location/switch off interfering device if possible) Should other error messages occur, switch balance off and then on again. If the error message remains inform manufacturer. KMA-TM-BA_IA-e-1310...
The addition of weighing scales capacity, preload and the total zero setting range give the required capacity for the weighing cell. To avoid overloading of the weighing cell, include an additional safety margin. • Smallest desired display division • Verifiable, if required KMA-TM-BA_IA-e-1310...
11.3 How to connect the platform Disconnect the display unit from the power supply. Solder the individual leads of the load cell cable onto the circuit board, see diagrams below. Weighing cell (Zemic L6D) Display Unit KERN KMA-TM KMA-TM-BA_IA-e-1310...
Confirm selected function by . The current setting will be displayed. Select desired setting by and confirm with , the balance returns to the menu. Exit menu/ Press , the balance will return to weighing mode. Return to weighing mode KMA-TM-BA_IA-e-1310...
1, div 2, div 5, div 10, Readability, s div 20, div 50 Balance capacity (/max) Adjustment, see chap. 16.1 P3 Pro tri* Not documented CoUnt Internal A/D converter value rESEt Reset weighing scale to factory setting SEtGrA Not documented * default setting KMA-TM-BA_IA-e-1310...
, the currently set position of the decimal dot is displayed. Press to select the desired setting. Selectable 0, 0.0, 0.00, 0.000, 0.0000. Confirm input with , the balance returns to the menu. Use to select the next menu item [inC]. KMA-TM-BA_IA-e-1310...
Page 25
Wait for stability display "STABLE", then confirm with The size of the currently set adjustment weight will be displayed. To change, select the digit to be altered by , and the (example) numerical value by Confirm with , [LoAd] appears. KMA-TM-BA_IA-e-1310...
Page 26
An adjusting error or incorrect adjusting weight will be indicated by the error message; repeat adjustment procedure. An adjusting error or incorrect adjustment weight will generate an error message („Err 4“), repeat the adjustment process. KMA-TM-BA_IA-e-1310...
The weighing result of the balance is outside the error limit. Therefore, in • regular intervals load balance with known test weight (ca. 1/3 of the max. load) and compare with displayed value. The reverification deadline has been exceeded. • KMA-TM-BA_IA-e-1310...
Page 28
Position adjustment switch and seals: Self-destroying seal mark Cover Adjustment switch KMA-TM-BA_IA-e-1310...
Prepare the required adjustment weight. The weight to be used depends on the capacity of the scale. Carry out adjustment as near as possible to the scale’s maximum weight. Info about test weights can be found on the Internet at: http://www.kern-sohn.com. • Observe stable environmental conditions. Stabilisation requires a certain warm-up time.
Page 30
The size of the currently set adjustment weight will be displayed. The active digit is flashing. To change, select the digit to be altered by , and the (example) numerical value by Confirm with , [LoAd] appears. KMA-TM-BA_IA-e-1310...
Page 31
An adjusting error or incorrect adjustment weight will generate an error message („Err 4“), repeat the adjustment process. 14 Enclosure Type Approval Certificate when using as weighing system KERN MCC or MPC KMA-TM-BA_IA-e-1310...