Dispensing calibration  Dispensing pump 1
V:
100 ml
t:
Press multifunction button  0/1
Stop the pump by pressing the 0/1 button, or it will stop automatically
after 60 seconds.
Measure the amount of liquid dispensed into the measuring vessel.
Dispensing calibration  Dispensing pump 1
V:
100 ml
t:
Enter the measured value into the "V:" field.
The throughput P is calculated in ml/min. by the controls.
Press the  "Save" button.
The measured value is allocated to the pump.
Reconnect the hose.

Cleaning and maintenance of the dispensing system

Rinse the dispensing system through with hot water every 2 weeks,
or before shutting the dispenser pumps down for any length of time,
to avoid the system getting clogged up and corroded.
To do this, rinse out the siphons with water.
Put the siphons in a container with warm water (104 - 122 °F / 40 -
50 °C).
Activate the dispenser calibration switch in each pump, until the
dispensing system is rinsed through.
Check the connections, hoses, dispensing sockets and seals for
leaks.
0 s
P: No
0/1
30 s
P: 200 ml/min
0/1