Dispensing
throughput (D1 to
D13)
Place the hose of the dispenser pump which is being checked at an
upward gradient in a measuring vessel.
 Irritant detergents.
Irritant for skin and eyes
Avoid contact with skin and eyes, wear protective clothing.
Return to the main menu by pressing  "Back".
Dispensing calibration
Dispensing pump 1
Dispensing pump 2
Select a dispenser pump.
Dispensing calibration  Dispensing pump 1
V:
100 ml
t:
Press multifunction button 0/1 .
Stop the pump by pressing the 0/1 button, or it will stop automatically
after 60 seconds.
Measure the amount of liquid dispensed into the measuring vessel.
Dispensing calibration  Dispensing pump 1
V:
100 ml
t:
Enter the measured value into the "V:" field.
The throughput P is calculated in ml/min. by the controls.
Press the  "Save" button.
The measured value is allocated to the pump.
Reconnect the hose.
The throughput of the connected dispensing pumps can be set from
0 ml/min to 9999 ml/min in 1 ml steps.
Call up the display as follows:
Supervisor level  Settings
Dispensing flow rate
Dispensing correction factor
Select "Dispensing throughput"
After selection, the following appears in the display:
Main menu
0 s
P: No
0/1
30 s
P: 200 ml/min
0/1
Programming
153