APPENDIX 1 – MESSAGES DISPLAYED
Message
AUTONOM:
CLEAR ALL? (Enter)
CLEAR LAST? (Enter)
Data saved
DATA SIZE:
HOLD
Password:
Invalid date
Energy Measuring
Memory Full
No ext supply!
No parameter sel
No Phase selected
PASSWORD ERROR
PASSWORD OK
Please wait
Recording
Too many param
Too many records
No Unit selected
ERR: SEQ
ERR: P-
ERR: SEQ & P-
ERR: CONNECTION
Error Vref
ERR: SYNC
Selection Error
Error1 - Error 5
Description
Available memory autonomy for the recording which is
being effected
The operator is trying to cancel all the recordings effected
The operator is trying to cancel the last recording effected
The data have been saved
Dimensions of the stored data
By pressing the correct key, the HOLD function has been
activated
A recording has been started and at least 5 minutes have
passed from the last activity of the instrument (see
paragraph 7).
The inserted date is not correct
The instrument is taking an energy measurement
The memory of the instrument is full
A recording has been started without connecting the
external power supply (optional code MAR####)
A recording has been started without selecting any value to
be recorded
Voltage and/or current harmonics have been selected and
the corresponding flag has been enabled (HARMONICS
ON) but no phase voltage or current has been selected
The inserted password is wrong (see paragraph 7).
The inserted password is correct
The instrument is waiting for the recording to be started
(see paragraph 6)
The instrument is recording (see paragraph 6)
More than 63 parameters have been selected (harmonics
included) or More than 38 parameters with CO-
GENERATION Flag enabled
The quantity of recorded data + Smp exceeds the
maximum allowed (35)
The Phase Sequence isn't correct.
The active powers shown on the right side of the message
are negative
The active powers shown on the right side of the message
are negative and the Phase Sequence isn't correct.
The instrument has detected a wrong connection to
Voltage inputs
The user set a Voltage reference not compatible with
voltage at instrument's input.
The System Frequency is out of range
There is a mismatch between the Parameter enabled and
the parameter selected for an AUX recording.
The instrument memory is damaged.
Page 129 of 131
Advices
Press ESC in order not to cancel the whole
memory, press ENTER to confirm
Press ESC in order not to cancel the last
recording, press ENTER to confirm
Press HOLD again to disable this function
Insert the password:
F1, F4, F3, F2
Check the inserted date
Press F1 to stop it
Cancel some recordings after transferring
them to a PC
Verify if you really want to start the recording
without the external power supply. In that
case press START again.
Press START/STOP and select at least a
value entering the MENU
Select at least one phase voltage and/or
current
Check the password
Deselect some values
Cancel some recordings after transferring
them to a PC
Check the Phase Sequence connection.
If there isn't a situation of co-generation
check if the clamps are properly connected
If there isn't a situation of co-generation
check if the clamps are properly connected /
check the Phase Sequence connection.
Check the Voltage connections
Check Voltage Reference set in "CONFIG
RECORDER"
Check the System Frequency, check setting
in ANALYZER CONFIG.
Check the parameter enabled in AUX
position and the selected parameter for
recording.
Contact Martindale Electric
VR2250