6.1.1. Calibrating the test leads ("CAL" Mode)

1. Connect the Red and Black test leads to B1 and B4 input terminals
respectively.
Connection of the test leads during calibration procedure.
2. If the test leads supplied with the instrument are not long enough for the
measurement you can extend the black cable.
3. Short-circuit the measuring cable ends making sure that the conductive parts of the
crocodiles make a good contact to each other (see previous picture).
4. Press the F2 key. The instrument carries out the calibration.
Never disconnect the test leads when the message "MEASURING" is
displayed.
5. At the end of the test the result is stored and used as OFFSET (that is to say
that it is subtracted from any continuity test carried out) for all the
subsequent measurements.
Note:
The instrument performs the calibration only if the resistance of the test leads is
lower than 5Ω.