6.2.4. "MAN" and "TIMER" mode faulty cases
If the instrument
detects the External
Power supply adapter
connected to the
instrument, the
message along side
will be displayed.
If the instrument
detects a Voltage
between Input
terminals higher than
15V, the instrument
does not perform the
test and displays the
screen alongside for 5
seconds.
If the instrument
cannot generate the
Nominal Test Voltage
it will emit a long
acoustic signal and
displays a screen
similar to the screen
alongside.
The displayed result can be stored pressing the SAVE key twice (according to
paragraph 9.1).
MΩ
05.06.01
Ω
-.- -
M
---V
15s
REMOVE POWER
MAN
500V
FUNC
VNOM
MΩ
05.06.01
Ω
-.- -
M
---V
15s
VOLT IN INPUT
MAN
500V
FUNC VNOM
These results cannot be saved
MΩ
05.06.01
Ω
1.17
M
107V
15s
MAN
500V
FUNC
VNOM
Page 24 of 131
Disconnect the External Power
Supply Adapter
ATTENTION: the test cannot be
executed. Check that the circuit is
not energised.
Insulation Resistance
ATTENTION: the test of
resistance R
was performed
ISO
at a voltage value lower than
the set rated voltage. Low
insulation case. This case
occurs under low insulation
conditions or in the presence of
capacity on the installation.
Test Time
VR2250