6.4.7. LOOP
Faulty Cases
If the instrument detects
the External Power
supply adapter
connected, the
instrument will show the
message displayed
along side.
Should the instrument
detect that the phase
and/or neutral cables
are not connected to an
installation, the screen
alongside is displayed
when pressing START.
Should the instrument
detect a voltage
between phase and
neutral higher of 250V,
for example in case the
black cable is connected
to an installation phase
conductor of a 400V
three-phase system, the
screen alongside is
displayed.
LOOP
05.06.01
Ω
- - -
---A
FRQ=50.0HZ
VP-N=231V
VP-PE=230V
REMOVE POWER
P-N
FUNC
Z2Ω
LOOP
05.06.01
Ω
- - -
---A
FRQ=50.0HZ
VP-N=
1V
VP-PE=
NO VOLTAGE
P-PE
FUNC
Z2Ω
LOOP
05.06.01
Ω
- - -
---A
Freq=50.0HZ
VP-N=401V
VP-PE=230V
HIGH VOLTAGE
P-PE
FUNC
Z2Ω
Page 45 of 131
Disconnect the External
Power Supply Adapter
0V
NO VOLTAGE Detected
50V
UL
HIGH VOLTAGE Detected
50V
UL
VR2250