6.3.3. RCD Faulty cases

6.3.3.1.
Connection troubles
If the instrument detects
the External Power
supply adapter
connected, the
instrument will show the
message displayed
along side.
Should the instrument
detect that the phase
and/or neutral cables
are not connected to an
installation, screen
alongside is displayed
when pressing START.
Should the instrument
detect a voltage
between phase and
neutral higher of 250V,
for example in case the
black cable is connected
to an installation phase
conductor of a 400V
three-phase system, the
screen alongside is
displayed.
RCD
05.06.01
- - -
ms
FRQ=50.0Hz
Ut= ---V
VP-N=230V
Vp-PE=230V
REMOVE POWER
x1
30mA
50V
FUNC
IdN
RCD
UL
RCD
05.06.01
- - -
ms
FRQ=50.0Hz
Ut= ---V
VP-N=
0V
Vp-PE=
LOW VOLTAGE
x1
30mA
50V
FUNC
IdN
RCD
UL
RCD
05.06.01
- - -
ms
FRQ=50.0Hz
Ut= ---V
VP-N=401V
VP-PE= 230V
HIGH VOLTAGE
x1
30mA
50V
FUNC
IdN
RCD
UL
Page 33 of 131
Disconnect the External
Power Supply Adapter
0V
NO VOLTAGE Detected
HIGH VOLTAGE Detected