Viewing Alarms; Rel Function - AEMC 6526 User Manual

Megohmmeter
Hide thumbs Also See for 6526:
Table of Contents

Advertisement

Insulation:
10V : <10kΩ, <20kΩ and <40kΩ
25V : <25kΩ, <50kΩ and <100kΩ
50V : <50kΩ, <100kΩ and <200kΩ
100V : <100kΩ, <200kΩ and <400kΩ
250V : <250kΩ, <500kΩ and <1MΩ
500V : <500kΩ, <1MΩ and <2MΩ
1000V : <1MΩ, <2MΩ and <4MΩ
In each measurement mode, the third threshold can be replaced by a user-defined value. To do this:
1. Press the
2. The > symbol starts blinking; you can change it to < by pressing the
of the alarm threshold: < for a low threshold and > for a high threshold.
3. To change the threshold setting, press the
change its value.
4. Use the
and
measurement.
5. When finished setting the threshold, press the

2.1.3.3 Viewing Alarms

When the measurement is below a low alarm threshold or above a high alarm threshold, the instrument emits a continuous
audible signal and the LCD indicates the threshold crossed:
ALARM
When checking continuity, this enables you to determine whether or not the continuity measurement is less than 2Ω simply
by listening, without looking at the display. You can similarly check insulation quality.
2.1.4 ∆REL Function
For an insulation, resistance, or capacitance measurement, you can configure the instrument to subtract a
reference value from the measured value and display the difference.
To activate this function, take a measurement, and then press the ∆REL button. This measurement becomes
the reference (Rref) and will be stored and subtracted from subsequent measurement values (Rmeas). The
REL
∆REL symbol appears on the LCD while this function is activated.
14
1.800.561.8187
button while the threshold value is displayed.
buttons to select and change the other digits in the threshold value, as well as the units of
<
www.
button to navigate to the first digit, and then use the
button to validate the setting.
ALARM
.com
button. This symbol indicates the direction
>
k
M
Megohmmeter Model 6526/6532/6534
information@itm.com
button to
M

Advertisement

Table of Contents
loading

This manual is also suitable for:

65326534

Table of Contents