Test Button Operation; Timed Tests - AEMC 6526 User Manual

Megohmmeter
Hide thumbs Also See for 6526:
Table of Contents

Advertisement

The
symbol indicates that the instrument is generating a hazardous voltage (>70V).
6. At the end of the measurement, release the TEST button. The instrument stops generating the test voltage and
discharges the device being tested. The
fallen below 70V.
Do not disconnect the leads and do not start any measurement while the symbol
When you release the TEST button, the measurement results remain displayed until the next measurement, or the HOLD
button is pressed, or the instrument is turned OFF.

2.3.2.1 TEST Button Operation

Pressing the TEST button starts an insulation measurement. In normal mode, the test voltage is generated for as long as
the button is pressed. When the button is released, the measurement stops.
In
mode, press the test button once to start the measurement, then press it a second time to stop; there is no need to
keep the button pressed. However, if you do not stop the measurement, it will stop automatically after 15 minutes.
In timed test mode (
the end of the defined test duration time.

2.3.2.2 Timed Tests

The TIMER
button activates timed test mode. This button is active only for insulation measurements.
1
press
st
2
press
nd
3
press
rd
PI T1T2
4
press
th
DARPI
5
press
th
Megohmmeter Model 6526/6532/6534
1.800.561.8187
> 1s
TEST
) press the TEST button once to start the measurement. The test will stop automatically at
, DAR, PI
This locks the TEST button. After you start the measurement, it continues to run without
requiring you to keep the TEST button pressed. The test will run until you stop it, or
when 15 minutes have passed.
This activates timed test mode. You can set a test duration between 1 and 39:59 min-
utes. Use the  and  buttons to modify the value displayed.
When the time duration is displayed, press the  button to enter edit mode. When the
first digit blinks, you can change it using the  button. Press  to go to the next digit
and  to change it. Then press  to validate.
This enables the PI function. This is used to calculate the polarization index (the ratio
of the measurement at 10 minutes to the measurement at 1 minute).
This enables the DAR function. This is used to calculate the dielectric absorption ratio
T1T2
(the ratio of the measurement at 1 minute to the measurement at 30 seconds).
Exits timed test mode.
www.
G
M
symbol is displayed until the voltage on the system under test has
.com
M
V
is displayed.
information@itm.com
19

Advertisement

Table of Contents
loading

This manual is also suitable for:

65326534

Table of Contents