Except when measuring voltage, all measurements must be made on powered-off systems. Therefore check to
ensure there is no voltage on the system under test before making a non-voltage measurement. When the rotary
switch is set to the voltage or an insulation testing position, the instrument measures and displays any voltage present
at the input terminals prior to the user pressing the test button.

2.1 Setting Up the Instrument

2.1.1 Configuration Settings

A >2 second press of the SET-UP button enables you to change configuration settings on the instrument.
You can then use the ◄ and ► buttons to scroll through and modify parameters.
> 2s
In Set-Up mode, the  button performs the following functions:
1
press
st
on 
2
press
nd
on 
3
press
rd
on 
4
press
th
on 
5
press
th
on 
6
press
th
on 
12
1.800.561.8187

2. OPERATION

The alarm buzzer is active. To deactivate it:
1. Press . On will blink to indicate it is selected.
2. Press  to change the setting to OFF.
3. Press  to validate the change.
The
symbol disappears from the display when you exit Set-Up.
Note that this setting reverts to On when you turn OFF the instrument.
Automatic switching to standby mode is activated. To deactivate it:
1. Press  to select OFF (the setting blinks).
2. Press  to change the setting to On.
3. Press  to validate the change.
P
The
symbol appears on the display when you exit Set-Up.
P
Note that this setting reverts to OFF when you turn OFF the instrument.
Displays the instrument model number.
Displays the instrument firmware version.
Displays the instrument hardware version.
Return to the first press.