This instrument is compliant with safety standard IEC 61010-2-030, and the leads are compliant with IEC 61010-031, for
voltages up to 600V in CAT IV or 1000V in CAT III.
Failure to observe the following safety instructions may result in electric shock, fire, explosion, and damage to the instrument
and installation.
Carefully read and understand all precautions for use.
Be aware of all electrical hazards when using this instrument.
Using this instrument other than as specified may compromise its user protection features.
The safety of any system in which this instrument is incorporated is the responsibility of the integrator of the
system.
This instrument can be used on CAT IV installations, for voltages not exceeding 600V
700V
maximum between terminals.
RMS
Do not use the instrument on networks whose voltage or category exceeds those specified in this manual.
Observe all environmental conditions of use (see § 3).
Except for voltage measurements, take no measurements on electrically "live" systems.
Do not use the instrument if it appears damaged, incomplete, or poorly closed.
Before each use, check the condition of the insulation on the leads, housing, and accessories. Any part on which
the insulation is deteriorated (even partially) must be set aside for repair or scrapping.
Using the instrument without its battery compartment cover may result in electric shock to the user.
Before using your instrument, ensure it is completely dry.
Use only the leads and accessories supplied. The use of leads (or other accessories) of a lower voltage rating or
category limits the use of the instrument/leads (or accessories) combination to the lowest category and service
voltage.
When handling the leads, test probes, and alligator clips, keep your fingers behind the physical guards.
Before removing the battery compartment cover, ensure all measurement leads and accessories are
disconnected. Replace all batteries at once. Use alkaline batteries.
Use personal protection equipment where appropriate.
All troubleshooting and metrological checks must be done by competent, accredited personnel.