2.1.2 Bluetooth
The instrument includes built-in Bluetooth class II capability. This enables you to connect the instrument to a computer and
then communicate with it via the DataView Megohmmeter Control Panel software (see § 4.3).
Connecting the instrument to a computer is a three-step process:
1. Enable Bluetooth on the instrument (§ 4.3.1).
2. Pair the instrument to the computer using Bluetooth (§ 4.3.2).
3. Connect the instrument to the computer through the DataView Megohmmeter Control Panel (§ 4.3.3).
2.1.3 Alarms
The instrument includes an alarm function that sounds an audible buzzer when a defined alarm condition is measured.
2.1.3.1 Activating/Deactivating the Alarm Function
The alarm is available in insulation, resistance, and continuity measurement modes. Pressing the
button activates the alarm. The
To turn OFF the alarm buzzer while it is sounding, press the HOLD button.
To deactivate an active alarm function, press the
2.1.3.2 Setting an Alarm Threshold
While
is displayed indicating the alarm function is active, you can change the alarm threshold by pressing the
(except during insulation measurements). For each testing mode, there are three pre-defined threshold values:
■
Continuity: < 2Ω, <1Ω and <0.5Ω
■
Resistance: >50kΩ, >100kΩ and >200kΩ
Megohmmeter Model 6526/6532/6534
1.800.561.8187
symbol is displayed, along with the threshold value.
<
ALARM
button.