CHAPTER 3 Settings and Measurement
(2)-3 Ring capture setting
(1) Ring Capture
(2) Number of Ring
Capture Points
(3) Ring Capture
Time
●Ring Capture Function
 Ring-captur has been operating in this instrument is as follows.
3−26
Setting
Sets the ring capture function to On or Off.
Specifies the number of data points in one file when the ring capture
function is On (See the following figure for details).
Displays the possible measurement time with one file when the ring
capture function is On.
Start capturing data
Captured data is saved into the File 1.
When the specified number of datas are saved into the File 1, the captured data is saved into the File 2.
The number of data is specified in the "Ring Capture Points" menu.
When the specified number of datas are saved into the File 2, the captured data is saved into the File 3.
The File1 is deleted at this point.
File1
File2
File3
When capturing is stopped at the STOP poinf in the above, the File4 and the File 5 are remained.
These files are consolidated into one file and it is saved. Then the ring catpuer is completed.
File4
File5
Twice as many files as the Number of Ring Capture Points will be created at the maximum.
Description
When the specified number of datas are saved into the File 3,
the captured data is saved into the File 4.
The File2 is deleted at this point.
When the specified number of datas are saved into the File 4,
the captured data is saved into the File 5.
The File3 is deleted at this point.
Repeat this action.
Stop capturing data
File4
File5
Filexx・・・: Saved