<Temperature Ranges>
Range
Maximum SPAN
K
-270 to +2000°C
J
-270 to +2000°C
T
-270 to +2000°C
R
-270 to +2000°C
E
-270 to +2000°C
B
-270 to +2000°C
S
-270 to +2000°C
N
-270 to +2000°C
W
-270 to +2000°C
PT100
-270 to +2000°C
JPt100
-270 to +2000°C
Pt1000
-270 to +2000°C
<Humidity Range>
Range
Maximum SPAN
(1)-3 Filter
Selection item
Off
No moving average is calculated.
2
A moving average is calculated twice per sampling interval.
5
A moving average is calculated 5 times per sampling interval.
10
A moving average is calculated 10 times per sampling interval.
20
A moving average is calculated 20 times per sampling interval.
40
A moving average is calculated 40 times per sampling interval.
<Filter processing>
Filter processing performed on the GL820 is the moving average shown in the following figure.
n-5
n-4
n-3
n-2
If the filter setting is 5,
D=(n-4 + n-3 + n-1 + n)÷5
If the sample interval exceeds 30 seconds, the average value of data obtained in a sub-sample (30
seconds) is used.
Minimum SPAN (p-p)
50°C
50°C
50°C
50°C
50°C
50°C
50°C
50°C
50°C
50°C
50°C
50°C
Minimum SPAN
(p-p)
0 to +110%
1.0%
Description
n-1
n
D
CHAPTER 3 Settings and Measurement
Measurement
Minimum Resolution
Range
-200 to +1370°C
-200 to +1100°C
-200 to +400°C
0 to +1600°C
-200 to +800°C
+600 to +1820°C
0 to +1760°C
0 to +1300°C
0 to +2000°C
-200 to +850°C
-200 to +500°C
-200 to +500°C
Minimum
Resolution
0.1%
0.1°C
3−19